5 Calibration
OI/ASO550–EN Rev. G
21
5 Calibration
This section describes how to calibrate the analyzer once it is
operational. Calibrations are initiated via the
Cal
prompt
displayed on
Operator
pages, or via the
Calibration Level
and
Advanced Level
menus or as a
Scheduled
calibration.
5.1 Accessing the Configuration level menus
The configuration level menus are used to configure the
wet-section(s) and set parameter values – see Fig. 7.1, page 42
for an overview of menus.
To enter
Configuration Level
menus:
1. Press the
key (below the
icon).
The
Operator
menus are displayed:
2. Press the
key to select the
Enter Configuration
menu
and press the
key (below the
icon).
The
Access Level
page is displayed:
3. Use the
key to scroll to the
Advanced
access level
and press the
key (below the
Select
prompt) to enter
the top level
Configuration
menus.
Use the
/
keys to scroll between top level menus
and press the
key (below the
Select
prompt on each
top level menu) to enter that level.
Note.
Calibration menus can be accessed from the
Advanced
level only – refer to Section 5.1 to access
the
Calibrate
level.
If calibration standard solution values need to be
changed from their default values, the values must be
set from the
Advanced
/
Calibrate / Calibration Setup
menu only – refer to Section 5.1 to access the
Calibrate
level.
Caution.
Do not calibrate the analyzer until the wet-section and
transmitter are installed and ready for operation – refer
to Section 4, page 19.
Allow at least 1 hour for the wet-section to stabilize
before running a calibration.
Note.
Refer to Section 7, page 42, for menu
descriptions.
Refer to Section 6.1, page 29, for details of menu
navigation and parameter selection / adjustment.
CAL
2012–04–15
10 : 34: 52
°C
21.7
21.7
2.93
Navigator
Tag 1
Na
ppb
2012–04–15
10 : 34: 52
°C
21.7
21.7
4.28
4.28
Operator Pages
Navigator
Data Views
Logs
Alarm Acknowledge
Manual Hold
Autoscroll
Media Card
Enter Configuration
ppb
Access Level
Calibrate
Advanced
Service
Back
Select
Read Only