Navigator 500
Sodium analyzer
3 Installation
OI/ASO550–EN Rev. G
9
3.1.3 Mounting the wet-section
Refer to Fig. 3.2 for wet-section dimensions. The wet-section weighs 4.5 kg (10 lb) – excluding bottle carrier and solutions.
Referring to Fig. 3.3:
1. Mark the wall using the dimensions shown.
2. Drill and plug 3 holes
A
and
B
in the wall suitable for
M6 or
1
/
4
in. fixings.
3. Screw in top fixing
A
, leaving a gap of 20 mm
(0.78 in.) between the fixing head and the wall.
4. Hang the wet-section onto fixing
A
, ensuring the
wet-section is retained firmly against the wall.
5. Secure the wet-section to the wall using 2 fixings
B
.
Note.
Clearance – the enclosure doors can open 180°. If mounting in a confined area, allow sufficient clearance for
door opening.
Dimensions in mm (in.)
Fig. 3.2 Sodium wet-section dimensions
Navigator 500
480 4
80
185
185
110.5
110.5
218.7
218.7
384.8 CTRS
Ø 7 (0.27)
2 Holes Ø 7
(0.27)
384.8
(15.15)
210 (8.27)
480
(18.90)
185 (7.28)
66
8
(26.
30)*
*When installing, allow a further 60 mm (2.36 in.) min. to remove the bottle carrier assembly,
or 135 mm (5.31 in.) min. to remove the calibration solution bottle from the carrier.
37 (1.45)
18
(0.7)
80.2
(3.15)
110.5
(4.35)
281
(11.06 in.)
290
(11.41 in.)
Note.
It is not possible to adjust fixing
A
once the
wet-section is placed over it.
Dimensions in mm (in.)
Fig. 3.3 Mounting the wet-section
386
(15.19)
210 (8.27)