Navigator 500
Sodium analyzer
7 Menu descriptions
OI/ASO550–EN Rev. G
47
Menu
Comment
Defaults
Solution Detection
Enables / disables the solution detection function.
Flow Meter
Note
. Single-stream wet-sections only.
Enables / disables flow measurement.
Disabled
Flow Meter 1 – 3
Note
. Multi-stream wet-sections only.
Enables / disables flow measurement for the selected stream.
Disabled
Stream Settings
Note
. Multi-stream wet-sections only.
Sample Time
Sets the how long each stream is sampled.
15 min.
Sequence
Sets the order in which the streams are sampled, a maximum of 8
entries are permitted. A 0 indicates the end of the sequence.
An example sequence for streams 1 to 3 is: 1,2,3,1,2,3,0 where the 3
streams are sampled consecutively.
1,2,3,1,2,3,1,2,1
(3-stream)
Stream 1 (3)
Enables / Disables the selected stream (disabled streams are not sampled).
Man. Valve Control
The solenoid valve can be energized manually. The current
concentration value is shown for reference. When the
Man. Valve
Control
parameter is exited, the valve returns to the closed state.
Note
: Press the
key to open the valve manually, press the
key
to close the valve manually.
Low Standard Valve
Enables the
Low Standard Valve
for manual control.
High Standard Valve
Enables the
High Standard Valve
for manual control.
Stream Valve
Note
. Single-stream wet-sections only.
Enables the
Stream Valve
for manual control.
Stream Valve 1 – 3
Note
. Multi-stream wet-sections only.
Enables the selected
Stream Valve
for manual control.
Regeneration Valve
If fitted, enables the
Regeneration Valve
for manual control.
Reset to Defaults
Resets all wet-section parameters to their default values.
Disabled