78 OI/FEP610/FEH610-EN Rev. B |
ProcessMaster FEP610, HygienicMaster FEH610
12
Specifications
NOTE
The detailed device data sheet is available in the download
area at www.abb.com/flow.
12.1
Permitted pipe vibration
In accordance with EN 60068-2-6.
Applicable to sensors in remote mount design and sensors in
integral mount design.
— Maximum deflection: 0.15 mm (0.006 inch) in the
10 ... 58 Hz range
— Maximum acceleration: 2 g, in the 58 ... 150 Hz range
12.2
ProcessMaster - Temperature data
Storage temperature range
-30 ... 70 °C (-22 ... 158 °F)
The temperature range offered by the device is dependent on
a number of different factors.
These factors include the measuring medium temperature
T
medium
, the ambient temperature T
amb.
, the operating
pressure P
medium
, the liner material and the approvals for the
explosion protection.
12.2.1
Maximum permissible cleaning temperature
CIP medium
Liner material Cleaning
temperature
Steam
PTFE, PFA
150 °C (302 °F)
Cleaning fluid
PTFE, PFA
140 °C (284 °F)
— The specified cleaning temperature applies for a maximum
ambient temperature of 25 °C (77 °F).
If the ambient temperature is > 25 °C (> 77 °F), the
difference to the actual ambient temperature must be
subtracted from the maximum cleaning temperature.
— The specified cleaning temperature may be applied for a
maximum of 60 minutes.