Request a mirror cleaning kit from ABB or your local distributor.
If further maintenance is required, contact ABB or your local distributor for
service.
o
ABB call center: 1-800-858-3847 (North America) or +1 418-877-2944
(Worldwide)
o
ABB Technical Support:
Other Components
The inline particulate filter requires replacement once a year or when the measured
data profile becomes thinner in the
Spectrum
display screen (GOF gets worse –
refer to the
section for details), and/or when the gas pressure
reading on the Control Bar increases.
Solid-state drive (SSD) maintenance: It is recommended to check the amount of
memory space available on the 120 GB SSD. Prior to collecting measurement
samples, verify there is sufficient SSD space available. In the Control Bar (see
Figure 66), the current amount of available disk space is displayed.
Figure 66: Current Disk Space Available
Should the analyzer shut down without a soft shutdown, the operating software
may get corrupted and make the analyzer non-functional. You can monitor the
internal lithium-ion battery energy level by checking the Control Bar. The lithium-
ion battery contains a protective PCB to prevent over-charging and over-discharge
to prevent damage to the battery. The lithium-ion battery takes approximately 8
hours to recharge completely. Inspect the lithium-ion battery for any bulge in its
package and protective PCB sealing. A bulge in the battery indicates internal
damage. Should the battery have a bulge, remove the battery and dispose it at an
organization that recycles lithium-ion batteries.
Battery Replacement
The analyzer battery may be replaced if it fails or is at end of life due to multiple
charge/discharge cycles.
CAUTION!
Always replace the battery only at an ESD protected station with
proper personal grounding. Any operation requiring opening the
analyzer case must be performed in a controlled environment with
proper ESD protection and the analyzer powered off. Never open
the analyzer without ESD protection.