ABB GLA131 Series Microportable Analyzers
User Manual
ABB PROPRIETARY INFORMATION
53
NOTE
Control of the MIU is unidirectional. The analyzer does not receive
feedback on the MIU state. If the MIU is enabled in the analyzer
Setup
panel, the data file is tagged with MIU valve descriptions
whether or not the MIU is properly connected. The data file simply
logs the condition of the control signal to the MIU.
Figure 46 shows the back panel of the eight-port MIU. The MIU inlet ports are
labeled numerically on the back panel of the MIU. The outlet port connects to the
gas inlet on the analyzer. The MIU is shipped with one each of these accessories:
Control cable (connects the analyzer to the MIU)
Power cable (powers the MIU)
¼ inch x 6 foot Teflon tube (connects the outlet port of the MIU to the inlet
port of the analyzer)
Figure 46: MIU Back Panel
Set Up the MIU
Connect the Components
1. Connect the provided power cable into the fused power-entry module on
the back panel.
2. Connect the control cable from the MIU to the
TO MIU
port on the back
panel of the analyzer. Figure 47 illustrates the pin numbers on the analyzer
DB9 male MIU output connector. Refer to Table 9 for pin numbers and
corresponding MIU port numbers.