![ABB GLA131 Series Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/abb/gla131-series/gla131-series_user-manual_3045448010.webp)
ABB GLA131 Series Microportable Analyzers
User Manual
ABB PROPRIETARY INFORMATION
10
There are one or two lasers (depending on your configuration) used in the analyzer.
During normal operation, with the analyzer case cover closed and the lock installed,
the analyzer spectroscopy instrument is a Class 1 Laser Product in accordance with
Title 21 Code of Federal Regulations, chapter 1, sub chapter J
.
The laser(s) in the analyzer are rated Class 3B, > 5 mW. The laser is enclosed and
not accessible, unless the lock is removed, and the case cover is opened for
servicing. Laser warning labels are on the enclosure covering the laser.
Figure 2: Laser Radiation Labels and Magnetic Interlock Switch
The analyzer laser is not field serviceable. Should a laser fail in the field, the
analyzer must be shipped back to ABB for service repair.
The analyzer laser is equipped with a magnetic interlock switch. Whenever the case
cover is open, the magnetic switch breaks contact from the magnetically encoded
counterpart, and power is interrupted from the Laser Controller PCB in the
analyzer, thus, disabling the laser.