The transmitted intensity also provides important information about the analyzer
measurement module light source. A drop in intensity may indicate an issue with
the laser, detector, ICOS module mirrors, or supporting electronics board. This
information is helpful when contacting ABB service about the issue.
Timechart Display
The
Timechart
display (see Figure 29) provides the absorption (in ppm or ppb) of
the sample gas measured. The black line represents the trace of measured
concentrations at a customizable interval rate.
Figure 29: Timechart Display
Refresh in the
Timechart
display (see Figure 30) refreshes the screen to display the
sampling data with a new starting point.