80
GAA330-M
EMISSION MONITORING SYSTEM | OI/GAA330-M-EN REV. C
Analyzer cabinet problems
Problem Cause Solution
Flow to analyzer not con-
stant / too low
Solenoid valve plugged
Untie sampling line on both ends of the sole-
noid valve and carefully clean the valve from
both sides with compressed air
End piece of sampling
probe black / plugged
Accumulation of hydrocar-
bons
Untie sampling line from solenoid valve before
cleaning the complete line with compressed air
from the top of the stack to the black / plugged
end
Humidity in one of the filter
elements
Gas cooler malfunctioning
Check function of the gas cooler. If not work-
ing, see above "Sample gas cooler problems"
Gas flow to analyzer dis-
rupted
Water trap closed after
contact with condensate
Replace the water trap
"Temperature Alarm" in
status signal board blinking
red
Heating in sampling probe
or sampling line damaged
Pt100 broken, needs to be replaced by service
personnel
Temperature controller
defective
Temperature controller needs to be replaced
by service personnel