EMISSION MONITORING SYSTEM | OI/GAA330-M-EN REV. C
Status messages
Information displayed on the ship control system
•
Calibration analyzer
•
Back-flush active
•
Collective error
•
Measuring range selection SO
2
•
MR1 0 to 250 ppm
•
MR2 0 to 500 ppm
Information displayed on the analyzer system
•
Temperature alarm
•
Fault cooler
•
Fault humidity
•
Flow alarm min.
•
Flow alarm max.
•
Condensate level
Sample gas cooler problems
Problem Cause Solution
No display
No power
Check power supply
Fuse blown
Check fuse and change it if necessary
Cooler not running
High temperature at the
compressor casing
Wait until cooled off and care for enough ven-
tilation
Status-LED blinks with high
temperature
Operational temperature
not yet reached
Wait for 20 minutes maximum
Cooling capacity to low, even
though cooler is running
Make sure that air can circulate free and that
ventilation louvers are not obstructed
Gas flow / dew point / gas
temperature too high
Check application parameters, install
pre-separator
Fan broken
Check fan, replace if necessary
Status-LED blinks with low
temperature
Control defect
Send cooler for inspection
Condensate in gas outlet
Condensate flask full
Drain flask
Stuck valve in automatic
condensate drain
Flush both directions
Cooler overloaded
Check limiting parameters
Reduced gas flow
Clogged gas path
Check / flush heat exchanger
Condensate outlet clogged
by ice
Send cooler for inspection
Display toggles between
temperature and error
message Error 1
Broken wire
Temperature sensor defect. Send cooler for
repair
Display toggles between
temperature and error
message Error 2
Short circuit
Temperature sensor defect. Send cooler for
repair