Electrical connection
50
FV4000-VT4/VR4 / FS4000-ST4/SR4
D184B097U02
5.3
M12 plug-in connector, PROFIBUS PA
As an option, the electrical connection can also be made using an M12 plug-in connector (see
the section titled "Ordering information"). The device can be shipped completely prewired. The
M12 plug-in connector is installed in the terminal box instead of the cable gland.
Important
Matching sockets (type EPG300) can be found in data sheet 10/63
−
6.44 xx, along with other
accessories.
G00653
4
1
3
2
Fig. 34: Assignment for connection using optional M12 plug-in connector (view from the front
looking at pin insert and pins)
Pin
Assignment
1 PA+
(31)
2 NC
3 PA-
(32)
4 Shield