52
FEP630, FEH630
ELECTROMAGNETIC FLOWMETER | CI/FEP630/FEH630-EN REV. D
… 6 Commissioning
… Parameterization via the menu function Easy Setup
Easy Setup
Pulse Width
30.00 ms
Next
Edit
22.
Use
to call up the edit mode.
23.
Use
/
to select the desired pulse width for the pulse
output..
24.
Confirm the selection with
.
Easy Setup
Damping
30.00 ms
Next
Edit
25.
Use
to call up the edit mode.
26.
Use
/
to set the desired damping.
27.
Confirm the selection with
.
Easy Setup
Dig.Out 41/42 Mode
Pulse
Next
Edit
28.
Use
to call up the edit mode.
29.
Use
/
to select the desired operating modeOff,
Logic, Pulse, Frequencyfor the digital output.
30.
Confirm the selection with
.
Easy Setup
Curr.Out at Alarm
High Alarm
Next
Edit
31.
Use
to call up the edit mode.
32.
Use
/
to select the desired alarm mode.
33.
Confirm the selection with
.
Easy Setup
Low Alarm
3.500 mA
Next
Edit
34.
Use
to call up the edit mode.
35.
Use
/
to set the desired current for Low Alarm.
36.
Confirm the selection with
.
Easy Setup
High Alarm
21.800 mA
Next
Edit
37.
Use
to call up the edit mode.
38.
Use
/
to set the desired current for High Alarm.
39.
Confirm the selection with
.