Endura APA592
pH / Redox (ORP) transmitter
2 Use in Areas Requiring Ignition Protection
12
OI/APA592–EN Rev. E
2.5 Hazardous Area Relevant Information
2.5.1 APA592-pH.A1… (Intrinsic Safety)
ATEX/IECEx approved for use in zone 0/20.
Ex ia (Zone 0):
LCIE 11 ATEX 3058 X
IECEx LCI 11.0050X
Designation
Device design: II 1 G Ex ia IIC T4
Housing design: II 1 D Ex iaD A20 IP66 T135 ºC, –20 ºC
T
amb
60 ºC
Safety-relevant data
Note.
The explosion-proof or ignition-proof designation is displayed on the agency certification label.
Input Parameters
Maximum voltage
U
i
= 30 V
Maximum input current
Ii = 160 mA
Maximum power
P
i
= 0,8 W
Internal inductance
L
i
= 0,5 mH
Internal capacitance
C
i
= 5 nF
Table 2.1 Intrinsic Safety Input Parameters
Output Parameters
Open-circuit voltage (maximum)
U
o
= 11.8 V
Short-circuit current (maximum)
I
o
= 5 mA
Maximum output power
P
o
= 15 mW
Allowed inductance (total)
L
a
= 1 H
Allowed capacitance (total)
C
a
= 1.45 µF
Table 2.2 Intrinsic Safety Output Parameters
Notes.
Parameters apply to entire system inclusive of cables.
Each specified electrical parameter must be applied individually and in combination. Do not
exceed the maximum values when applying the electrical parameters individually or in
combination.
Both Intrinsic safety and dust ratings are combined when APA592-pH.A1… is ordered.