Endura APA592
pH / Redox (ORP) transmitter
6 Configuration
OI/APA592–EN Rev. E
45
6 Configuration
6.1 Configuration Options
APA592–PH transmitter configuration options comprise:
Configuration via the transmitter's navigation keys.
Configuration via FDT/DTM technology.
Configuration can be performed with any FDT network applications that are approved for use with
the DTM (for example, ABB AssetVision Basic / Professional). The bus can be connected via FSK
modem as well as HART + USB or HART Multiplexer.
Configuration via EDD technology.
Configuration can also be performed with EDD master applications such as Siemens Simatic PDM,
that is approved for use with EDD.
Configuration using a handheld terminal.
Note.
Complete configuration of the transmitter is available through the HMI. However, not all
Device DTM functions (for example, such as Diagnosis Masking) are accessible via the
transmitter's navigation keys.
Communication with the transmitter is supported by the HART protocol – refer to Section
9, page 83 for HART commands.
Fig. 6.1 Configuration Options – Schematic
APA592–PH Transmitter
Handheld
Device
FTD/DTM or EDD
Technology
Modem
ABB Asset Vision
or DCS Config