7 Operator Pages and Menus
60
OI/APA592–EN Rev. E
7
At the calibration results screen(s) check
the new values.
Press the
key to exit
the screen:
or
if a calibration failure message is
displayed, reset the buffer values (see
steps 2 to 6).
8
If the
Hold Output
screen was selected at
step 1, an option to release the output is
provided at the
Release Output
screen.
Press the
key to retain the
Hold
Output
state or the
key
release the
Hold Output
state.
Step
Procedure
Screen
Table 7.3 2-Point Manual Calibration (Continued)