Endura APA592
pH / Redox (ORP) transmitter
8 Troubleshooting and Diagnostics
78
OI/APA592–EN Rev. E
S070.042
Snsr temperature
under range
Sensor temperature is below range.
[PROCESS]
Verify sensor wiring.
Verify configuration related to temperature sensor
and temperature compensation.
S072.041
Snsr temperature
over range
Sensor temperature is above range.
[PROCESS]
Verify sensor wiring.
Verify configuration related to temperature sensor
and temperature compensation.
S074.001
Temperature input error
Temperature related sensor board
electronics error.
[ELECTRONICS]
Check temperature sensor connections. If a
temperature sensor is not connected, switch
temperature compensation to
Manual
mode.
If the problem persists, contact factory.
S076.010
PV out of range limits
The analog output for the Primary
Variable is beyond its scaling limit
and no longer represents the true
applied process.
[PROCESS]
Adjust the working range if possible.
S078.004
PV out of physical limit
The process applied to the sensor
for the Primary Variable is beyond
the operating limits of the device.
[PROCESS]
Verify sensor wiring.
Verify configuration setting for the sensor.
Diagnostic Message
[Sub-error Message]
Possible Cause(s)
[Related Condition]
Corrective Action
Table 8.2 Diagnostic Messages (Continued)