42/24-30 EN Rev. 1
EL6010 Gas Analyzer in Category 2G Operator’s Manual
47
Status Messages,
continued
Nr. Status Problem
Reaction / Solution
12 M
Temperature
compensation
measured value exceeds the ana-
log/digital converter value range.
Check the plug connections in the gas analyzer. Call for
service.
13 M
Pressure
compensation
measured
value exceeds the analog/digital
converter value range.
Check the plug connections in the control unit. Check
the sample gas pressure. Call for service.
14
M
Temperature controller measured
value exceeds the analog/digital
converter value range.
Check the plug connections in the control unit. Call for
service.
15
M
Temperature above or below
upper and/or lower limit 1.
16
M
Temperature above or below
upper and/or lower limit 2.
Status messages during the warm-up phase. If the
status message appears after warm-up: Check if the
permissible ambient temperature range is being main-
tained (see page 8). Check the connecting leads and
plug connections. Check the line seating in the insu-
lated jackets. Check the thermal link and replace as
necessary.
20 M
F
Internal processor bus fault.
21 M
F
Internal display bus fault.
22 M
F
Internal I/O bus fault.
Switch power supply off and on again. Call for service.
30 M
Broken analog output line.
Check the analog output lines.
No.
Number of status message displayed in error list
Status
M
Maintenance Request: Status is displayed on the screen and output via Modbus. The gas
analyzer is in a state that will soon require user intervention. The measured value is valid.
F
Failure: Status is displayed on the screen and output on “overall status” digital output and
via Modbus. The gas analyzer is in a state that immediately requires user intervention. The
measured value is invalid.
Maintenance Mode: Status is output during calibration on “overall status” digital output and
via Modbus.
Содержание EL6010
Страница 1: ...Continuous Gas Analyzers EL6010 Operator s Manual 42 24 30 EN Rev 1 ...
Страница 2: ......
Страница 3: ...Continuous Gas Analyzers EL6010 Operator s Manual Publication No 42 24 30 EN Rev 1 Edition 07 04 ...
Страница 66: ...64 EL6010 Gas Analyzer in Category 2G Operator s Manual 42 24 30 EN Rev 1 Declaration of Conformity continued ...