EL6010 Gas Analyzer in Category 2G Operator’s Manual
42/24-30 EN Rev. 1
Chapter 1
Preparation for Installation
Installation Location Requirements
Installation Location
The gas analyzer should not be mounted outdoors.
Short Gas Paths
Install the gas analyzer as close as possible to the sampling location.
Locate the gas conditioning and calibration assemblies as close as possible to the
gas analyzer.
Adequate
Air Circulation
Provide for adequate natural air circulation around the gas analyzer. Avoid heat
buildup.
Protection from
Adverse Conditions
Protect the gas analyzer from:
•
Cold
•
Direct sunlight and heat
•
Large temperature variations
•
Strong air currents
•
Accumulations of dust and dust infiltration
•
Corrosive atmospheres
•
Vibration
Air pressure range
600 to 1250 hPa
Relative humidity
max. 75 %
during storage and shipping
–25 to +65 °C
EL6010-Caldos17
+5 to +50 °C
EL6010-Magnos106
+5 to +50 °C
Environmental
Conditions
Ambient temperature
in operation
EL6010-Uras14
+5 to +45 °C
Note: The explosion protection is not impaired if the gas analyzer is operated at
temperatures less than +5 °C and down to –20 °C. However in this temperature
range the compliance with the metrological data cannot be guaranteed.