EL6010 Gas Analyzer in Category 2G Operator’s Manual
42/24-30 EN Rev. 1
Calibrate Sample Component,
continued
Step Action
Key Display
Key
9
×
Ø
Span calibration
CO
Enter
Select
Span calibration
with test gas
.
10
×
Ø
Span setpoint
10.00
Vol% CO
Enter
Check the span set
point and adjust as
needed.
11
Connect span gas!
Enter
Turn on the span gas
supply.
12
Meas. value stable?
9.91 Vol% CO
Enter
Start calibration if the
measured value reading
is stable.
13
Wait cal. is running
Calibration in progress.
14
Accept calibration
or
Meas. value ok?
10.00 Vol% CO
Enter
reject
calibration.
Back
9
×
Ø
Span with cal. cell
CO
Enter
Select
Span calibration
with calibration cell
.
10
Setpoint cal. cell
8.52 Vol% CO
Enter
The calibration cell set
point is displayed.
11
Connect zero gas!
Enter
Connect the zero gas (if
this has not already been
done).
12
Meas. value stable?
8.49 Vol% CO
Enter
Start calibration if the
measured value reading
is stable.
13
Wait cal. is running
Calibration in progress.
14
Accept calibration
or
Meas. value ok?
8.52 Vol% CO
Enter
Option for
EL6010-Uras14
reject
calibration.
Back
Amplification error
0.00 Vol% CO
This display shows that an error occurred
during calibration and that calibration
was unsuccessful.
In this case, press
Enter
to return to the
start of zero or span calibration, check the
test gas concentration, correct as needed
and repeat the calibration process.