42/24-30 EN Rev. 1
EL6010 Gas Analyzer in Category 2G Operator’s Manual
55
EL6010-Uras14: Metrological Data
Sample Components and
Smallest Measurement Ranges
Sample
Com
ponent
Smallest
Class 1 Range
Smallest
Class 2 Range
Smallest Class 2
Range with
Calibration Cell
Gas
Group
1)
CO
0– 100 ppm 0–
10 ppm
0–
50 ppm
2)
A
CO
2
0– 100 ppm 0–
5 ppm
0–
25 ppm
2)
A
NO
0– 200 ppm 0– 150 ppm
0– 150 ppm
2)
A
SO
2
0– 100 ppm 0–
25 ppm
0–
25 ppm
2)
A
N
2
O
0– 100 ppm 0–
20 ppm
0–
50 ppm
2)
A
NH
3
0– 500 ppm 0–
30 ppm
–
B
CH
4
0– 100 ppm 0–
50 ppm
0–
50 ppm
2)
A
C
2
H
2
0– 200 ppm 0– 100 ppm
0– 100 ppm
B
C
2
H
4
0– 500
ppm 0– 300
ppm 0– 300
ppm
B
C
2
H
6
0– 100 ppm 0–
50 ppm
0–
50 ppm
2)
B
C
3
H
6
0– 250 ppm 0– 100 ppm
0– 100 ppm
2)
B
C
3
H
8
0– 100 ppm 0–
50 ppm
0–
50 ppm
2)
B
C
4
H
10
0– 100 ppm 0–
50 ppm
0–
50 ppm
2)
B
C
6
H
14
0– 500 ppm 0– 100 ppm
0– 100 ppm
2)
B
SF
6
0– 2000 ppm 0– 1900 ppm
0– 2000 ppm
B
H
2
O
0– 1000 ppm 0– 500 ppm
0– 500 ppm
C
1) See price information
2) The sample component is calibrated in the factoryset
measurement range which can be increased by a factor of 5.
Other sample components on request.
Stability
Linearity Deviation
≤
1 % of span
Repeatability
≤
0.5 % of span
Zero Drift
≤
1 % of span per week.
For ranges smaller than Class 1 to Class 2:
≤
3 % of span per
week
Sensitivity Drift
≤
1 % of measured value per week
Output Signal Variations
≤
0.2 % of span at 2
σ
and electronic T90 time = 5 sec (Class 1)
Detection Limit
≤
0.5 % of span
Measurement Range Data
Quantity
1 range per sample component.
The measurement range is freely adjustable within a range ratio
of
≤
1: 5 relative to the factoryset measurement range.
Largest Measurement Range
0 to 100 Vol.% or 0 Vol.% to saturation or 0 Vol.% to LEL.
Measurement ranges within ignition limits cannot be provided.
Measurement Ranges with Suppressed Zero-Point
Differential measurement with flowing reference gas.
Max. suppression ratio of 1:10 depending on the base level.
Influence Effects
Flow Effect
Flow rate in the 20–100 l/h range: 0.5 % of span at a flow rate
change of
±
10 l/h
Pressure Drop at Flame Barriers
Sample Gas or Reference Gas Inlet: Approx. 40 hPa at a flow
rate of 50 l/h. Purge gas inlet: Approx. 20 hPa at a flow rate of 10
l/h
Associated Gas Effect / Cross Sensitivity
Analyzer calibration should be based on an analysis of the sam
ple gas. Selectivity measures to reduce associated gas effect
(optional): Incorporation of interference filters, filter vessels or
internal electronic crosssensitivity correction or carrier gas cor
rection for a sample component by other sample components
measured with the analyzer unit.
Temperature Effect
Ambient temperature in permissible range
–
At zeropoint:
≤
1 % of span per 10 °C;
for ranges smaller than Class 1 to Class 2:
≤
2 % of span per 10 °C
–
On sensitivity with temperature compensation:
≤
3 % of measured value per 10 °C
–
On sensitivity with thermostat effect at 55 °C (optional):
≤
1 % of measured value per 10 °C
Air Pressure Effect
On sensitivity with pressure correction by means of integral
pressure sensor:
≤
0.2 % of measured value per 1 % barometric
pressure change.
Pressure sensor working range: p
abs
= 600–1250 hPa
Dynamic Response
Warm-Up Time
Approx. 30 minutes without thermostat; approx. 2 hours with
thermostat.
T
90
Time
T
90
= 2.5 sec for measurement cell length = 200 mm and sample
gas flow = 60 l/h without signal damping (low pass filter).
Materials in Contact with the Sample Medium
Analyzer (Sample Cells)
Tubing: Aluminum or goldplated aluminum; Window: CaF
2
(Option: BaF
2
); Connectors: Rust and acidresistant steel 1.4571
Gas Lines and Connectors
Rust and acidresistant steel 1.4571