50
User Guide
Updating with a USB key
NOTICE
If the update process do not seem to work on your first try, try it again with a different
brand of USB key or update remotely (see “Updating with a web browser (preferred
method)” on page 48).
1
From your computer, extract the zip file sent to you by the service representative at the root level of a
USB key.
2
Open the CoreSense M10 cabinet.
3
Shutdown the instrument using the main power switch.
—
Figure 53
Instrument main power switch
4
Insert the USB key on the USB port of the CoreSense M10 cabinet.
—
Figure 54
Instrument USB port
5
Turn on the instrument using the main power switch (see Figure 4). The update process will start
automatically and take a few minutes.
Messages will appear on the local screen during the update process. Also, the instrument may
reboot two or three times depending on the required updates for your system. The update process
is complete once a message appears indicating to remove the USB key and reboot.
6
When this message appears, remove the USB key.
7
Power off the system for a few seconds and power it on again.
You may be asked to calibrate the screen by touching the 4 crosses displayed on the screen.
Содержание CoreSense M10
Страница 1: ... USER GUIDE CoreSense M10 Multi gas monitoring system ...
Страница 8: ...Page intentionally left blank ...
Страница 18: ...Page intentionally left blank ...
Страница 24: ...Page intentionally left blank ...
Страница 26: ...22 User Guide Figure 19 Selecting an input channel ...
Страница 36: ...Page intentionally left blank ...
Страница 44: ...Page intentionally left blank ...
Страница 56: ...52 User Guide ...
Страница 62: ...Page intentionally left blank ...
Страница 66: ...Page intentionally left blank ...
Страница 107: ......