ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
… 12 Diagnosis / Troubleshooting
… Possible status messages
No.
Text
S A W F Reaction/Remark
116
The Profibus Module is mounted on the wrong
slot! The Profibus interface is therefore not
working. Please remount the Profibus Module on
slot X20/X21.
x x
See message text
117
The configuration backup was saved.
118
The configuration backup was loaded. The
system has been restarted.
119
The system configuration could not be loaded!
This system therefore contains no configuration
now. Please enter menu: Configure/System/Save
Save configuration to load your backup
configuration. Or use SMT to re-install your
configuration.
x x
See message text
System bus
201
The selected system bus module could not be
found.
x x
Check plug connections and terminating resistors on the system bus. Make sure the
system bus module serial number is correctly entered:
‘MENU / Diagnostic/Information / System overview’
203
The selected system bus module does not exist.
x x
Check plug connections and terminating resistors on the system bus.
208
The system bus was not able to transfer data into
the database.
x x
The system bus module software version is not compatible with that of the system
controller; update the system controller software.
209
The system bus connection
to this module is interrupted.
x x
Check the system bus connection to the indicated system bus module. Check the
power supply system of the indicated system bus module.
210
The system bus module configuration has
changed.
x x
For information; the configuration data are automatically updated
211
The system bus module has no more on-board
memory.
x x
Check the system bus module configuration:
‘MENU / Diagnostic/Information / System overview’
214
The system is currently maintained with Optima
SMT.
215
The analyzer module has an internal
communication error!
x x
Notify Service.
216
The analyzer module has an internal program
error!
x x
Notify Service.
250
The selected analyzer module could not be found! x x
Check the connectors and cabling.
251
The connection to the analyzer has been lost!
x x
Check the connectors and cabling.
252
The EEPROM data of the analyzer is faulty!
x x
Check the configuration with TCT.
253
Communication with the analyzer is faulty!
x x
Check the connectors and cabling.
254
The boot program of the analyzer is defective!
Notify Service!
x x
Notify Service.
255
The program of the analyzer is defective! Notify
Service!
x x
Notify Service.