30
AW T4 20 |
U N I V ER S A L 4 - W I R E , D UA L- I N P U T T R A N SM I T T ER | C I/AW T4 2 0 - EN R E V. B
…10
Sensor setup
pH/Redox/ORP
Menu
Comment
Default
Tag
Enter an alphanumeric sensor tag (16 characters maximum) to identify the sensor on the
Operator Pages
.
TAG1
Measurement Type
Select measurement type:
• pH/Redox/ORP
Note
. If a change is made the I/O sources are reset.
pH
Range High
Set the span value used in
Chart
and
Bargraph
views.
14.00
Range Low
Set the zero value used in
Chart
and
Bargraph
views.
0.00
Filter Type
Select the signal filtering type:
• None/Low/Medium/High
None
Note .
The following menus are displayed only if
Measurement Type
=
pH.
Temp . Compensation
Set the type of temperature compensation:
• Manual/Automatic/Auto solution
Automatic
Solution Coeff .
Note
. Displayed only if
Temp. Compensation
type =
Auto solution.
Set the solution coefficient (pH or mV change per 10 deg C) of the solution being monitored.
See Appendix A,
.
N/A
Manual Temperature
Note .
Displayed only if
Temp. Compensation
type =
Manual.
Enter the temperature of the sample within the range –10.0 to 120.0 °C.
N/A
Note .
The following menus are displayed only if
Measurement Type
=
Redox/ORP.
Temperature Sensor
Set the type of temperature measurement:
• Manual/Automatic
Note .
If
Temperature Sensor
type =
Manual
, the temperature value is not displayed in the associated
Operator page
or
Signals View
.
N/A
Low Slope Limit
A pH probe degrades over time. As this happens the slope calculated by a calibration procedure gradually
decreases. Set the slope value below which a calibration fails. The low slope warning diagnostic is
activated if the calibration calculates a slope less than 20 % above this value.
40%
Sensor Diagnostics
Broken Glass
Note:
Displayed only if
Measurement Type
=
pH.
Detect broken glass condition:
•
Enabled/Disabled
Disabled
Out Of Solution
Detect Out Of Solution condition:
•
Enabled/Disabled
Disabled
Ref . Poisoning
Note:
Displayed only if a digital (EZLink) sensor is connected.
Detect a contaminated reference electrode:
•
Enabled/Disabled
Disabled
Ref . Failure
Note:
Displayed only if a digital (EZLink) sensor is connected.
Detect a failed reference electrode:
•
Enabled/Disabled
Disabled
Ref . Blocked
Detect a blocked reference electrode:
•
Enabled/Disabled
Disabled
Ref . Alarm Limit
Note:
Displayed only if
Ref. Blocked
sensor diagnostic is
Enabled.
A blocked reference electrode is detected when the impedance of the reference electrode exceeds
a given limit. Set the impedance value above which the reference blocked diagnostic is activated.
N/A
Reset To Defaults
Select to reset all
Sensor Setup
parameters to their default values.