  U N I V ER S A L 4 - W I R E , D UA L- I N P U T T R A N SM I T T ER  |  C I/AW T4 2 0 - EN R E V. B

33

Connectivity/Communications (optional)

Ethernet

HTTP, HTTPS, FTP, Secure FTP

PROFIBUS DP

DPV0, DPV1

MODBUS

RTU RS485

HART

• 

Fieldcomm certified version – HART 7

•  Configured range

 

4 to 20 mA, user-programmable across  
measurement range

• 

Dynamic range

 

3.8 to 20.5 mA with 3.6 mA low alarm level,  
21 mA high alarm level 

•  Accuracy 

 

±0.25 % of reading

•  Maximum load resistance

 

– 500 Ω at 20 mA 

•  Configuration 

 

Can be assigned to either measured variable

•  Isolation

 

– 500 V DC from any other circuitry

 

Data logging

Storage

•  Measurement value storage  

(programmable sample rate)

• 

Audit log*, Alarm log*, Calibration log, Diagnostics log

Storage media

SD card, up to 32 GB capacity

Chart view

On local display

Historical review

Of data

Data transfer

SD card interface – Windows-compatible FAT file system, 

 

data and log files in Excel and DataManager Pro compatible 

formats 

Environmental data

Ambient operating temperature:

–10 to 55 °C (14 to 131 °F)

Ambient operating humidity:

Up to 95 % RH non-condensing

Storage temperature:

–20 to 70 °C (–4 to 158 °F)

Altitude:

2000 m (6562 ft) max. above sea level

*  Audit log and Alarm log data are stored in the same log file.

2-electrode conductivity

Conductivity input

Measurement range and resolution

Cell 
constant

Conductivity  

range

Display  

resolution

Accuracy 

repeatability

0.01

0 to 200 µS/cm

0.001 µS/cm

±1.0 % of 

measurement 

range per decade

0.05

0 to 1000 µS/cm

0.001 µS/cm

0.1

0 to 2000 µS/cm

0.01 µS/cm

1

0 to 20000 µS/cm

0.1 µS/cm

Dynamic response

<3 s for 90 % step change when damping is Off

Damping

Configurable: Off, Low, Medium and High

Temperature input

Temperature element types

•  Automatic temperature sensor recognition for Pt100, 

Pt1000 and 3k Balco RTDs in either 2-lead or 3-lead 

configurations

•  Temperature element can be used for automatic 

temperature compensation of the conductivity solution

Measurement range and resolution

Sensor 
group

Temperature  

range

Display  

resolution

Accuracy 

repeatability

Pt100 
Pt1000

 

3K Balco

–20 to 200 °C 

 

(–4 to 392 °F)

0.1 °C 

 

(0.1 °F)

0.1 °C

 

(0.18 °F)

None

User-programmable  

–20 to 300 °C 

 

(–4 to 572 °F)

N/A

Temperature compensation modes

Linear, UPW, NaCl, HCl and NH

3

Reference temperature

25 °C (77 °F)

Configured output range

Cell constant

Min . span

Max . span

0.01

1 µS/cm

200 µS/cm

0.05

5 µS/cm

1000 µS/cm

0.1

10 µS/cm

2000 µS/cm

1

100 µS/cm

20000 µS/cm

Содержание AWT420

Страница 1: ...nance procedures refer to the specific sensor manual For more information Further publications for the AWT420 transmitter are available for free download from www abb com measurement or by scanning th...

Страница 2: ...trode sensors 12 pH and conductivity connections 13 pH ORP pIon sensor module connections 13 Power supply connection 14 Fitting the EZLink modules 16 Connecting EZLink sensors 18 Long cables 18 6 Easy...

Страница 3: ...nk PCB upgrade spares kit 37 Communications module assemblies 37 AWT420 HART PCB upgrade spares kit 37 AWT420 Profibus PCB upgrade spares kit 37 AWT420 Modbus PCB upgrade spares kit 37 AWT420 Ethernet...

Страница 4: ...ury To ensure safe use when operating this equipment the following points must be observed Up to 240 V AC may be present Be sure to isolate the supply before removing the terminal cover Safety advice...

Страница 5: ...ed down if it has been installed to IP66 NEMA 4X standards i e cable glands are correctly fitted and all unused cable entry holes are blanked off see page 11 and page 10 Warm water and a mild detergen...

Страница 6: ...igure 1 AWT420 transmitter main components 4 Mechanical installation Transmitter installation Optional accessories Optional installation accessories Cable gland kit Weathershield Panel mount kit Pipe...

Страница 7: ...CE If the optional weathershield D is used position it between the transmitter and wall and pass 2 screws C through fixing holes both sides in weathershield Referring to Figure 5 1 Position the left a...

Страница 8: ...The correct torque is critical to ensure proper compression of the panel seal and achieve the IP66 NEMA 4X hosedown rating 4 Holding assembly F together position bracket D into the left hand recess on...

Страница 9: ...e holes to suit vertical or horizontal pipe secure the clamp plate to the pipe mounting bracket B using two M6 x 8 mm hexagon head screws and spring lock washers C 3 Position the pipe mounting bracket...

Страница 10: ...Profibus and Ethernet communication wiring ONLY A special cable gland is supplied with the Ethernet communications option and should be used only for the Ethernet cable The use of cable glands cable...

Страница 11: ...s module PCB slot Communication module connections refer to separate supplements HART COM AWT420 HART EN MODBUS COM AWT420 MODBUS EN PROFIBUS COM AWT420 PROFIBUS EN ETHERNET COM AWT420 ETHERNET EN AC...

Страница 12: ...Red Black Brown Green Yellow Blue 2085 2 lead Red Blue Yellow Dark green TB2 2 lead Green Black Blue Yellow Dark green AC2xx 2 lead Green Black Blue Red Yellow Dark green 3 lead Green Black Yellow Red...

Страница 13: ...d Clear Black Red White TB5 2 lead Blue Black Red White AP1xx 2 lead Clear Black Red Red White 3 lead Clear Black White Red Red AP3xx 2 lead Blue Black Red White 3 lead Blue Black Red White Grey APS1x...

Страница 14: ...ly 1 cable per cable gland Referring to Figure 11 page 15 1 Using a suitable screwdriver release door retaining screw A and open the transmitter door 2 Release terminal cover retaining screw B and rem...

Страница 15: ...420 EN REV B 15 Figure 11 Connecting the transmitter AC mains power supply A D E F G H 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 GLASS GUARD REFERENCE SOLUTION EA R TH PT PT SCREEN PT 3 rd COMMUNICATIONS Sensor...

Страница 16: ...onnector for sensor 1 passes through cable entry D b if two EZLink modules are used push fit sensor 1 module into location C and sensor 2 module into location E Note When fitting the cable assemblies...

Страница 17: ...nnector body using the flats on the alignment washer see Figure 13 9 Screw back nut J onto connector body and tighten using a spanner Figure 14 Securing the EZLink connector cable assembly Referring t...

Страница 18: ...Setup prompt is displayed Press the key to start Easy Setup or press the key to cancel and exit to the main Operator page Press the key Edit to change the default value setting to the required value s...

Страница 19: ...items and decrease displayed values E Group key Toggles between Operator pages 1 to 5 when an Operator page is selected with the View key View screens Diagnostics View Signals View Alarms View and Out...

Страница 20: ...g CAL shortcut D Press the key C below the CAL prompt Figure 19 Operator menus Operator menus comprise Operator Pages displays the Operator page for each available sensor Data Views displays enabled d...

Страница 21: ...H and turbidity The highest priority diagnostic or alarm is displayed Other active diagnostic alarm states can be viewed on the Diagnostics View see page 22 Figure 20 Operator Page multiple sensors 4...

Страница 22: ...fer to the sensor s Operating Instruction the bargraph flashes to indicate the value is outside the specified range When multiple sensors are fitted and Autoscroll is selected from the Operator Menu s...

Страница 23: ...View 01 O2 O3 O4 CAL 04 09 2019 08 10 11 Diagnostics View PV Failure NV Error Cal Failed Cleaning NAMUR icon and message priority see page 64 Diagnostic message View icon No Message CAL 04 09 2019 08...

Страница 24: ...shown in Table 3 The Diagnostics Log shows the history of diagnostic messages that have been displayed in the Diagnostic View see page 26 Log entry Description Cal Failed Calibration procedure failed...

Страница 25: ...toring Do not expose to static electricity electrical noise or magnetic fields When handling an SD card take care not to touch any exposed metal contacts Back up critical data stored on removable medi...

Страница 26: ...29 Access level screen Level Access Logout Displayed only after Calibrate or Advanced levels are accessed Logs the user out of the current level If passwords are set a password must be entered to acce...

Страница 27: ...A Range High View the span value used in Chart and Bargraph views N A Range Low View the zero value used in Chart and Bargraph views N A Filter Type Select the signal filtering type None Low Medium Hi...

Страница 28: ...gs Note The temperature compensation characteristic TC Curve for signal A should be set to NaCl Note The temperature compensation characteristic TC Curve for signal B should be set to NH3 Difference C...

Страница 29: ...string 6 characters maximum for the custom user defined concentration units N A Conc Curve Table Set the user defined concentration curve using the 6 point linearizer table concentration against condu...

Страница 30: ...surement Manual Automatic Note If Temperature Sensor type Manual the temperature value is not displayed in the associated Operator page or Signals View N A Low Slope Limit A pH probe degrades over tim...

Страница 31: ...4 mbar 380 to 835 mm Hg If the barometric pressure is unknown leave at the default sea level value of 1013 mbar 760 mm Hg 1013 mbar Reset To Defaults Select to reset all Sensor Setup parameters to the...

Страница 32: ...it hubs or EZLink connectors Security Password protection Access to configuration levels is enabled only after the user has entered a password Calibrate level user assigned password Advanced level use...

Страница 33: ...4 to 158 F Altitude 2000 m 6562 ft max above sea level Audit log and Alarm log data are stored in the same log file 2 electrode conductivity Conductivity input Measurement range and resolution Cell c...

Страница 34: ...eference temperature 25 C 77 F Configured output range Sensor group Min span Max span A 100 S cm 2000 mS cm B 10 S cm 2000 S cm pH ORP Redox pH ORP Redox input Sensor types pH Glass Antimony Sb ORP Re...

Страница 35: ...s with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference recieved including...

Страница 36: ...cause interference and This device must accept any interference including interference that may cause undesired operation of the device Le pr sent appareil est conforme aux CNR d Industrie Canada app...

Страница 37: ...spares kit Part number 3KXA877420L0011 EZLink module assemblies AWT420 EZLink PCB upgrade spares kit Part number 3KXA877420L0015 Communications module assemblies AWT420 HART PCB upgrade spares kit Pa...

Страница 38: ...A877210L0103 Weathershield and pipe mount kit Part number 3KXA877210L0104 Gland packs EZLink connectors Gland packs Part number 3KXA877420L0111 3KXA877420L0112 3KXA877420L0113 3KXA877420L0114 3KXA8774...

Страница 39: ...oid is a trademark of Google LLC Bluetooth is a registered trademark of Bluetooth SIG Inc HART is a registered trademark of the FieldComm Group iOS is a trademark of Apple Inc registered in the U S an...

Страница 40: ...this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior...

Отзывы: