42/24-12-2 EN
Advance Optima Ex Analyzer Modules Operator’s Manual
13
Connect Electrical Lines - Instructions,
continued
Isolation Relay -K01
Isolation relay -K01 is installed in the CU-Ex. It is used to connect the system bus
cable HIGH, LOW and GROUND leads and the non-intrinsically safe system
controller signal leads.
The standard arrangement of isolation relay -K01 is shown in the CU-Ex Operator’s
Manual (Publication No. 42/24-13 EN).
Terminal Strip -K05
Terminal strip -K05 is installed in the CU-Ex. It is used to connect the 8-wire
system bus cable leads not connected to isolation relay -K01.
Terminal strip -K05 is not needed if the 8-wire Ex analyzer module system bus
cable is routed to a connection box and a 3-wire extension cable is used for the
connection to the CU-Ex.
Terminal Strip -K06
Terminal strip -K06 is installed as an option in the central unit version for use in
Zone 2 and in the non-explosion protected version. It is used for system bus cable
connections.
If terminal strip -K06 is not installed, the Ex analyzer module system bus cable
should be routed to a connection box and the connection to the central unit
should be made via the standard system bus cable with plug and a tee.
Power Line Filter -Z01
The Ex analyzer module 24-VDC connecting cable can only be connected directly
to the (Ex) central unit if the optional -Z01 power line filter is installed.
If the -Z01 power line filter is not installed, the 24-VDC cable is connected to a
separate power supply.
F
For further information see the CU-Ex Operator’s Manual (Publication No. 42/24-13
EN).