Control commands (SPECTRAN:CTRL)
The SPECTRAN control commands are used to read or to write parameter of the device, like Startfrequency,
Stopfrequency RBW, etc.
They are different for the SPECTRAN HF and SPECTRAN NF devices. Therfore the overview is divided into the
common part for all devices, and a HF / NF part with the additional device parameter.
Foreach SPECTRAN:CTRL command you will receive two ACMD packages as answer. Please see declaration
to that effekt in the topic
Protocol types
, section ACMD - Spectran control command protocol)
Overview (common parameter):
SPECTRAN:CTRL:SWEEPING [ ? | value] (or SPEC:CTRL:SWEEP)
Description: Check if device is measuring, or start / stop measuring
Command parameter:
? read state
1 enable measurement
0 disable measurement
Answer:
ACMD
Package with message
SPECTRAN:CTRL:STARTFRQ [ ? | value] (or SPEC:CTRL:START)
Description: Read or write Startfrequency
Command parameter:
? read Startfrequency
value write Startfrequency (value must be in the frequencyrange of the analyzer)
Answer:
ACMD
Package with message
SPECTRAN:CTRL:STOPFRQ [ ? | value] (or SPEC:CTRL:STOP)
Description: Read or write Stopfrequency
Command parameter:
? read Stopfrequency
value write Stopfrequency (value must be in the frequencyrange of the analyzer)
Answer:
ACMD
Package with message
SPECTRAN:CTRL:CENTFRQ [ ? | value] (or SPEC:CTRL:CENT)
Description: Read or write Centerfrequency
Command parameter:
? read Centerfrequency
value write Centerfrequency (value must be in the frequencyrange of the analyzer)
Answer:
ACMD
Package with message
SPECTRAN:CTRL:SPAN [ ? | value] (or SPEC:CTRL:SPAN)
Description: Read or write Spanfrequency
Command parameter:
? read Spanfrequency
value write Spanfrequency (value must be in the frequencyrange of the analyzer)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание Spectran V4
Страница 26: ......
Страница 34: ...System Overview This chapter describes all system overview pages...
Страница 35: ...Spectran Server Overview Networkserver Settings Timestamp Device information...
Страница 38: ...List of clients Shows a list of currently connected clients One row per client...
Страница 40: ...Phone Add a contact phone number Contact Add a contact person...
Страница 41: ...Server Logfile Shows information of the default logfile...
Страница 46: ...System Configuration This chapter describes all system configuration pages...
Страница 50: ...Main menu Main program menu...
Страница 63: ...Save Information...
Страница 66: ......
Страница 67: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 68: ......
Страница 69: ...Save Information button...
Страница 70: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 75: ...Delete a selected profile from RSA This can t be undone...
Страница 84: ...The file list shows all recorded files in the selected folder Folder list...
Страница 92: ...Help Information This page contains useful information and links At the moment you ll find only placeholder...
Страница 105: ......