background image

Operating instructions for moisture analysers 
UX 3081 
UX 3081WQ 

 
a&p instruments, Kluckhofer Weg 61, D-32657 Lemgo, Tel.:05232 97780, Fax.: 977820  
E-Mail: [email protected], Internet: www.apinstruments.de 

Seite 12 

Saving drying parameters in the 

program memory

If your device has a program memory, appears in the top left of the display
  "QP" for type UX3081 and UX3081WQ and a number of1-9.

depending on which heater the program memory was activated for from the manufacturer, 
here for the infra-red heater.

time

mode

value (weight)

unit

UX 3081WQ

g

0,000

A

QP1

The standard parameters are set fom the manufacturer at the program memory place 1 .

The different drying parameters can be saved or be changed with button         .
Drying parameters are: temperature and the parameters of the automatic or the timer.

P2

mode

value (weight)

unit

UX 3081

g

0,000

A

QP2

P2

set

up

down

mode

mode

value (weight)

value (weight)

unit

unit

Memory

Menory

automatic 2

automatic 1

Here you can choose between the 9 memory places.
Then the menu set the temperature, the weight loss, the query interval and the query start.
All other parameters and settings are not saved individually to the memory place.

set

set

set

set

set

set

set

set

set

set

1)

esc

1)

esc

1)

esc

up

down

up

down

up

down

up

down

down

down

down

down

down

time

time

time

time

time

mode

mode

mode

mode

mode

value (weight)

value (weight)

value (weight)

value (weight)

value (weight)

unit

unit

unit

unit

unit

Menu  HP2

Menu  HP2

Automatic

Automatic

Automatic

Temperature

Auto/Timer IR

Weight loss mg

Interval sec.

Start after min.

time

time

time

time

time

mode

mode

mode

mode

mode

value (weight)

value (weight)

value (weight)

value (weight)

value (weight)

Ceramikh.  

200

Automatic 

30 mg 

20 s 

10 min 

Temperature

Auto/Timer IR

Weight loss mg

Interval sec.

Start after min.

unit

unit

unit

unit

unit

--C

4 x

Changes for the quartz heater to the unit of measurement, printer settings, data output, device ID,  date and time,
Buzzer and language, same as temperature and the parameters of the automatic system affect all prog ram 
memories at the same time.

Содержание UX 3081

Страница 1: ...measurement time gentle and even drying of samples high reproducibility Automatic or time controlled measurement shut off UX 3081 3081WQ Special device for extra large samples e g in the chipboard ind...

Страница 2: ...tenance 17 Changing the print ribbon 5 Communication 18 Consumables 14 Data interface 14 Date and time 10 Display 6 Error description and solution 16 Features 1 ID number 9 Installation site 4 Intende...

Страница 3: ...ter Repairs should only be carried out by a qualified technician Please take extra care when using flammable samples You may need to contact the manufacturer for advice Please be careful not to drop a...

Страница 4: ...ion site Please check the packaging and device for any damage upon delivery Please retain the original packaging in case you need to return the goods to a p instruments Please only use the original pa...

Страница 5: ...on on the back of the device Once the device is switched on the current software version is shown followed by the weight display 1 7 In order to prevent weighing system measuring errors the device sho...

Страница 6: ...accessed using the up or down button In this instruction manual it is assumed that the down button follows the menu item sequence The menu is operated using a standard structure UP or DOWN Set parame...

Страница 7: ...ngs can be defined here 1 Weight loss in mg a value should be defined here which corresponds to approx 0 1 of the original sample e g original sample 10 g 10 mg When the weight loss the mg weight spec...

Страница 8: ...described for radiant heaters Measurement unit During measurement the device optionally displays moisture dry matter TS Example 50 moisture 100 Atro moisture Atro i e moisture related to the dry matt...

Страница 9: ...t can be selected separately down up down up down up set set set esc down set esc down esc set down up set esc down set set down esc temp C time mode value weight unit temp C time mode value weight un...

Страница 10: ...cales Tray carrier and tray must be on the balance Follow the menu instructions and use a F1 weight to calibrate the scales Tray being used must be empty If the display shows switch the power switch o...

Страница 11: ...ed The menu can be locked or unlocked as shown below Press and hold the set button for 5 secs Position Enter 5 digit 5x Under scored pin etc oder Position Enter 5 digit 5x Under scored pin etc If you...

Страница 12: ...tomatic 2 automatic 1 Here you can choose between the 9 memory places Then the menu set the temperature the weight loss the query interval and the query start All other parameters and settings are not...

Страница 13: ...re that the heater is swung back to its rearmost position 3 Measurement values are displayed for the entire length of timethe heater is positioned above the sample The device should not be moved from...

Страница 14: ...nd creates a COM interface as soon as the computer is connected to the moisture analyser If this is not the case the driver must be manually installed in the device manager Special software is require...

Страница 15: ...set Menu buzzer Buzzer activated deactivated Menu date time Date Day Month Year Time Hours Minutes Men ID Device ID 9999 Device ID 0000 Menu data output Data output without LF with LF Menu printer Pr...

Страница 16: ...tact a p instruments Menu Drying parameters cannot be entered only Menu is locked Unlock menu lock contact the person responsible or a p instruments Menu lock cannot be unlocked 5 digit pin for unlock...

Страница 17: ...Check shut off settings Timer settings incorrect Printer No printout Printer is faulty or Contact a p instruments no data transmission Contact a p instruments No linefeed Linefeed button or printer is...

Страница 18: ...er adjustable from 40 C to 180 C 3081 105 C to 360 C 3081QW Measuring units moisture dry matter ATRO moisture g solid matter kg Reading accuracy 0 1 readable Communication Data terminal V24 RS 232 USB...

Страница 19: ...Tray tongs 10000230 for lifting the tray supplied Mains cable H1000019 Mains connection supplied Calibration weight F1 200 g 10000403 for calibrating the scales Aluminium foil 130 mm x 260 mm H100001...

Страница 20: ...rating instructions for moisture analysers UX 3081 UX 3081WQ a p instruments Kluckhofer Weg 61 D 32657 Lemgo Tel 05232 97780 Fax 977820 E Mail info apinstruments de Internet www apinstruments de Seite...

Отзывы: