Operating instructions for moisture analysers
UX 3081
UX 3081WQ
a&p instruments, Kluckhofer Weg 61, D-32657 Lemgo, Tel.:05232 97780, Fax.: 977820
E-Mail: [email protected], Internet: www.apinstruments.de
Seite 5
1. 4. Next place one of the two drying trays supplied on
the tray holder using one or two sets of tongs.
The notch in the reflector tray acts as an access
gap for easy positioning of the tray.
1. 5. Check that the voltage indicated on the type plate matches the local mains voltage.
1. 6. The power cable is inserted into the
device’s power jack and a suitable
earthed mains supply. After the power
supply has been connected, the
device can be switched on or off using
the on/ off button on the back of the device.
Once the device is switched on, the current
software version is shown, followed by the
weight display.
1. 7. In order to prevent weighing system measuring errors, the device should be switched on 20 minutes
before being used to allow for heat compensation. If weight values vary in the display’s milligram range
during this period, this is considered normal within this phase. If the device is switched off using the reset
(on/ off) button, it does not require a warm-up period, if switched on again with the same button.
1 8. ULTRA X moisture measuring devices are supplied, so that in most cases substances can be measured
immediately. If you are not satisfied with the result, the measurement parameters can be adjusted:
Printer
Open Cover:
Change Paper
Line feed
Hold the green glowing putton
and lift the plastic cover
The ULTRA X Thermo paper is stable for 10 years.