GB 27
English
Calibration report example for weighing sensor compliant with GLP, GMP and ISO
For the AD-8121 printer, use printer MODE 3.
11.3. Calibration of Drying Temperature for MX-50
– The temperature calibrator (accessory AX-MX-43)
adjusts the drying temperature on the pan.
Put the sensor on the pan and input measurement
data at 100°C and 160°C.
– Each adjustment needs fifteen minutes.
The buzzer sounds at the end.
–
t-Up
is displayed after no adjustment for five
minutes during the operation and calibration is
stopped. Press any key to return to weighing
mode.
– Refer to the instruction manual of the certified
temperature calibrator (accessory AX-MX-43).
CALIBRATED
WEIGHT
CAL.WEIGHT
20.000 g
SIGNATURE
- - - - - - - -
A&D
MODEL
S/N
ID
DATE
TIME
MX-50K
K1234567
LAB-123
2002/04/15
13:57:24
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration mass
Signature
Sensor pan
Thermometer