AND MX-50/MF-50 Instruction Manual
GB 12
5.3. Self-Test Function (Motion Check)
Use the self-test function to check functionality and accuracy of the analyser.
5.3.1. Operation
1. Turn on the analyser.
The display will show 0.000 g.
2. Press and hold the PROGRAM key until
CH
appears.
3. Place the breeze break ring, pan support, pan handle
and sample pan in order. (Leave sample pan empty)
Close the heater cover.
Press the ENTER to start the test.
Example:
CLOSE
is displayed.
"Heater cover is open. Close the cover to start the test."
4. Normal test needs approximately one minute.
Good result:
CH pa55
is displayed few seconds and
buzzer sounds. The analyser returns to weighing mode
automatically.
Error:
Buzzer is beeped out.
An error code is displayed.
Example:
CH no
,
error 0
or
Ht err
0.000
0.000
PROGRAM
CH
CH . . .
ENTER
CH pa55
6.1. Standard Mode
The Standard Mode provides measurements of moisture content, simply by selecting desirable accuracy
and drying temprature. All other conditions are pre-set in the Analyser.
Accuracy
The accuracy of measurement can be selected from HI, MID. or LO. Sample mass and final drying rate
are automatically set by ACCURACY.
In the Standard Mode, the measurement will be completed when the Drying Rate become less than the
value set for ACCURACY.
Value and conditions set for each ACCURACY.
ACCURACY
Use
Sample mass
Drying Rate to Terminate measurement
MX-50
MF-50
HI
Precise result
10 g
0.02 %/min
0.05 %/min
MID
5 g
0.05 %/min
0.10 %/min
LO
Quick measurement
1 g
0.10 %/min
0.50 %/min
Note: There is a user mode that can set the measurement parameters in detail.
Refer to "7.Selection of Measurement Method".
M e a s u r e m e n t P r o c e d u r e
6