MXMF.Users.Handbook.V2.20.doc

LIT:MXUHB April 2003

4

-

-

2. Method of measurement

How moisture content is measured with a moisture analyzer

Moisture content can be measured by heating method, Karl Fischer method, Dielectric

?

Infrared absorption 

method, neutron analyzer or crystal oscillation method, among which mainly heating and drying method or 
Karl Fischer method is utilized in many laboratories. Infrared absorption method and Dielectric method are 
suitable for process.

Heating method is a method where moisture content is rendered by the weight of 

water evaporated from ground solid or liquid samples after heating the sample for a period of time at or 
over the sample’s perspiration temperature. Weight loss occurs as samples get heated and finally
converge on a certain value. Sample may pyrolyze and vaporize depending on its characteristics, which 
suggests that what is vaporized is not necessarily water. However suitable sample size and heating
temperature such as temperature setting, heating time and so forth can lead a result comparable to the 
one determined by Karl Fischer method.

Among all, moisture analyzers employed with heating method is easy to handle with its good operability
and clear formula of operation, low profile and can be run and maintained at a low cost, which contribute to 
attracting a wide range of users who test many kinds of samples. Measurement range is from 0.01% or 
0.1% to 100% and so samples even with almost 100% moisture content can be correctly and easily
measured. Heating method moisture analyzers utilizes halogen, infrared lamp, sheathed heater or micro
wave heater as heating source to heat sample on an electric balance that weighs the sample before and 
after the heating to determine the moisture loss. The electric balance then needs technology that insulates 
heat to the load sensor and prevents temperature drift since temperature can get over 150 – 200? . Karl 
Fischer method is a method where moisture content is measured by electric chemical titration under favor 
of the characteristics that in the presence of methyl alcohol, KF reagent including iodine, hydrogen
disulfide, pyridine responds specifically with water.

H

2

O + I

2

+SO

2

+3RN+CH

3

OH

?

2RN

?

HI+RN

?

HSO

4

CH

3

(anhydrate)

Water (reagent)

KF reagent

methyl alcohol

(

RN: base

?

I: Iodine

?

SO

2

: hydrogen disulfide

)

Содержание MX-50

Страница 1: ...yz ze er r U Us se er rs s H Ha an nd db bo oo ok k Version 2 20 April 2003 A D Weighing USA Version T T Te e es s st t t w w wi i it t th h h T T Th h he e e B B Be e es s st t t T T Tr r ru u ue e...

Страница 2: ...ment 4 3 Accuracy 6 4 Samples 6 5 Calibration 7 6 Others 8 7 Heating method 9 8 Measurement 9 9 Data Analysis 10 B Data Results Analysis 1 Measurement samples 11 2 Typical measurement results by table...

Страница 3: ...ce Card Power Cable 2 Fuses Specifications are subject to change for improvement without notice Standard 85mm 400W straight halogen lamp heating system with SRA filter and SHS weighing technology Halo...

Страница 4: ...ials are measured with a stable moisture content rate and a typical material with such character is Sodium Tartrate Dihydrate Known by its name it is a well known material secondarily produced while a...

Страница 5: ...e to the one determined by Karl Fischer method Among all moisture analyzers employed with heating method is easy to handle with its good operability and clear formula of operation low profile and can...

Страница 6: ...od doesn t however it is suitable for detecting moisture content in a material such as vapor or that with very little moisture in ppm Moisture analyzers using infrared absorption method work in the be...

Страница 7: ...rate is when chemically bonded to two water molecules or crystal water and becomes Sodium Tartrate again by releasing the water molecules when heated Its moisture content rate is the ratio of the two...

Страница 8: ...atable 9 Does placement of sample such as powder on a pan affect measurement result Yes How to put a sample on a pan determines heat distribution To measure moisture content rate accurately even heati...

Страница 9: ...t remove moisture in a sheet dry and store it in a desiccator prior to use 2 Is absolute measurement of moisture content rate possible Can only water content be measured No Minerals such as metal glas...

Страница 10: ...ble to see inside the analyzer while heating Being able to see inside of the analyzer gives users a sense of safety Not only water in a sample that can be overheated or is burnable may be vaporized bu...

Страница 11: ...aps like burns 9 Data Analysis No Q A 1 Can soybeans or coffee beans be tested as they are NO Usually samples like peas should first be crushed in a blender or mill because temperature of surface and...

Страница 12: ...0065 2 13 19 1 By the data there is no evident difference between the two analyzer s results and it could be said that data taken by A D s MX MF can match a Karl Fischer type analyzer 2 As for accurac...

Страница 13: ...Bond 6 Commodity Paste 1 Standard MID 200 9 7 61 3 0 309 0 5 Place the sample evenly on the pan Hand Soap 7 Commodity Liquid 1 Standard MID 200 6 92 01 0 157 0 17 Use a glass fiber sheet 8 Commodity...

Страница 14: ...140 8 5 3 04 0 01 0 33 Crush with hand mixer 29 Food Butter Peanuts 5 Standard MID 140 9 6 2 1 0 077 3 67 Crush with hand mixer Ground Coffee Beans 30 Food Powder 5 Standard MID 140 9 8 4 43 0 036 0 8...

Страница 15: ...051 1 07 47 Food Powder Horse Radish 5 Standard MID 140 11 4 3 7 0 082 2 22 Horse Radish 48 Food Grated 1 Standard MID 200 15 1 39 07 0 123 0 32 Crush the sample from above a glass fiber sheet Ginger...

Страница 16: ...rmicelli 2 Standard MID 200 15 8 14 8 0 15 1 01 Cut the sample by about 3cm 63 Food Dried Brown Seaweed 1 Standard MID 200 9 2 11 49 0 367 3 19 Crush with hand mixer Wood Ear 64 Food Sliced 2 Standard...

Страница 17: ...383 0 47 Use a glass fiber sheet 84 Food YoghurtB 1 Automatic 0 5 min 180 5 4 88 07 0 209 0 24 Measure after putting the sample between folded glass fiber sheet and pressing it hard 85 Food Soy Milk 1...

Страница 18: ...owder 1 Standard MID 200 2 5 11 24 0 591 5 26 104 Industrial Products Activated charcoal Particulate For deodrant use 5 Standard MID 120 6 6 9 96 0 142 1 43 Silica gel 105 Industrial Products Particul...

Страница 19: ...oftware WinCT Moisture WinCT Moisture in combined use with heating and drying method moisture analyzer MX 50 and MF 50 manufactured by A D enables easy transfer of measurement data to PC through RS 23...

Страница 20: ...ulated from decrease in sample quantity caused by vaporization of moisture from heating It is determined at a point where changes no longer occur displayed on graph with planarized line Measurement re...

Страница 21: ...mperatures due to the sample s high heat resistance Measurement of such samples can be completed in a short time by heating at the highest possible temperature Samples with similar measurement process...

Страница 22: ...where moisture content is stabilized and rapid changes in curve do not occur When stabilized moisture content curve re inclines it is assumed that materials other than water additives organic matter...

Страница 23: ...the same conditions sample quantity heating temperature measurement termination condition with previously set heating temperature and measurement termination condition terminates measurement when the...

Страница 24: ...value described with the vertical axis at the left Change of inclination could be seen in change of heating temperature Curve displayed in green plots inclination of moisture content curve red by 1min...

Страница 25: ...tent rate curve or primary derivative value Rate min RsTemp is a software designed to determine heating temperature most suitable for the sample from measured and calculated results However it is impo...

Страница 26: ...different temperatures The top graph displays results of measurement on starch below on cornstarch The Rate min constantly keeps low values at a higher temperature zone Measurement of such samples cou...

Страница 27: ...ncrease again Suspected causes to this increase of Rate above 180C is that either ingredients other than water liquid additives organic matter have vaporized or samples have started to carbonize In su...

Страница 28: ...measurement with heat and dry method moisture analyzer However it is possible to measure and evaluate moisture content of such samples by measuring under same conditions sample quantity heating temper...

Страница 29: ...ned glass housing affect moisture content rate measurement YES it may Through stained glass heat conduction may not be good enough to heat sample evenly which can lead to low repeatability One of the...

Отзывы: