![A&D GC Series Скачать руководство пользователя страница 70](http://html.mh-extra.com/html/aandd/gc-series/gc-series_instruction-manual_446421070.webp)
70
18.2.3.
Interface for Remote Scale
Available with the AD-8561-MI05.
With a load cell type of weighing platform, a two-scale system can be structured.
Sensitivity Adjustment of Remote Scale
When connecting a remote scale for the first time, set the weighing capacity and minimum display and perform
sensitivity adjustment using a weight.
1. Connect a remote scale via the AD-8561-MI05.
Prepare a weight (separately sold) in advance.
Connect the power 30 minutes or more before starting sensitivity adjustment. Turn on the display
(warm up).
2. Remove the cable panel and press the CAL switch. “
Cal
” is shown.
To cancel the operation, press the RESET key.
3. Press the REMOTE SCALE key to light up the remote scale
mark (remote scale is selected). Press the ENTER key.
4. Use the numeric keys 1 , 2 , 5 , and
・
to
enter minimum display. Press the ENTER key.
5. Use the numeric keys 0 to 9 and
・
to
enter weighing capacity. Press the ENTER
key.
The figure to the right shows an example of setting
weighing capacity to 60.00 kg and minimum display
to 0.01 kg.
6. “
” is shown. Weighing capacity/minimum display has been set and the remote scale is ready for
sensitivity adjustment.
7. Check that the remote scale mark is lit (remote scale is selected). Refer to “
” to perform sensitivity adjustment of the remote scale.
8. Press the REMOTE SCALE key in weighing mode to switch to the remote scale. Confirm that it can
properly weigh.
Remote scale mark
Weighing capacity
Minimum display