BM series
29
6.4.
Percent Mode (%)
The percent mode displays the weight value in percentage compared to a 100% reference mass
and is used for target weighing or checking the sample variance.
Selecting the percent mode
1 Press
the
MODE
key to select the unit
%
(Percent mode).
If the percent mode cannot be selected, refer to "
5. Weighing Units
".
Storing the 100% reference mass
2 Press
the
RANGE
key to enter the 100% reference mass
storing mode.
3 Place a container (tare) on the weighing pan, if necessary.
Press the
RE-ZERO
key to cancel the weight (tare).
The balance displays
100
0
%
.
4 Place the sample to be set as the 100% reference mass on
the pan or in the container.
5 Press
the
key to store the reference mass.
The balance displays
100.00
%
.
Notes
The decimal point position depends on the 100% mass.
BM-252
,
BM-500
BM-300
,
BM-200
BM-20
,
BM-22
100% mass
Minimum
display
100% mass
Minimum
display
0.0100 g to 0.0999 g 1 %
0.00100 g to 0.0099 g 1 %
0.1000 g to 0.9999 g 0.1 %
0.01000 g to 0.0999 g 0.1 %
1.0000 g to
0.01 %
0.10000 g to
0.01 %
If the balance judges that the mass of the sample is too light
(under 0.01g) to be used as a reference, it displays
lo
.
The 100% reference mass can be stored in the non-volatile
memory and is maintained even if the AC adapter is removed.
6 Remove the sample.
Reading the percentage
7 Place a sample to be compared to the reference mass on
the pan. The displayed percentage is based on the 100%
reference mass.
100% mass
Place sample
Percentage
Weighing pan
Container
(tare
)
Содержание BM series
Страница 1: ...Analytical Balance BM 20 BM 22 BM 252 BM 200 BM 300 BM 500 1WMPD4002195D ...
Страница 8: ...6 BM series ...
Страница 113: ...BM series 111 MEMO ...
Страница 114: ...112 BM series MEMO ...
Страница 115: ...13 ...