66
BM
series
Storing the weighing data
Note If
"
Clr
" blinks in the upper left of the display, delete the data in memory.
1 Set the "Data memory (
data
)" parameter to "
1
" (To store the weighing data).
2 Specify whether or not to add the time and date using the "Time/Date output (
5-td
)" parameter.
3 The storing mode depends on the "Data output mode (
prt
)" parameter setting.
Four modes are available to store data.
Key
mode ........................... When the
key is pressed and the displayed value is stable, the
balance stores the weighing data.
Auto print mode A.............. When the displayed value is stable and the conditions of "Auto print
polarity", "Auto print difference" and zero point (reference value) are met,
the balance stores the weighing data.
Auto print mode B.............. When the displayed value is stable and the conditions of "Auto print
polarity", "Auto print difference" and last stable data (reference value)
are met, the balance stores the weighing data.
Interval
mode...................... Weighing data is stored at an interval specified in "Interval time (
int
)".
Press the
key to start and stop this mode.
Enabling the data memory function
1 Press and hold the
RANGE
key (for approx. 2 seconds)
until
ba5fnc
is displayed, and then release the key.
2 Press
the
RANGE
key several times to display
dout
.
3 Press
the
key.
4 Press
the
RANGE
key several times to display
data
.
5 Press
the
RE-ZERO
key to display
data
1
.
6 Press
the
key to store the setting.
7 Press
the
CAL
key to return to the weighing mode.
Press and hold
(approx. 2 seconds)
Press several times
Press several times
Содержание BM series
Страница 1: ...Analytical Balance BM 20 BM 22 BM 252 BM 200 BM 300 BM 500 1WMPD4002195D ...
Страница 8: ...6 BM series ...
Страница 113: ...BM series 111 MEMO ...
Страница 114: ...112 BM series MEMO ...
Страница 115: ...13 ...