BCD OUTPUT (AD-8922A-01)
Outputs the weighing data received from the weighing instrument in BCD format, along with the polarity
(+/-) and the data status (stable/unstable and over/under).
Using the STROBE signal, the data can be read easily. The AD-8922A-01 can set the input terminal
function to either the BUSY or the RE-ZERO input function by selecting that function. When set to the
RE-ZERO input function, the AD-8922A transmits the RE-ZERO command to the weighing instrument,
setting the weighing value to zero. BUSY input enables the data to be held or prevents data refreshing
during the reading operation.
The logic of data, status and strobe can be switched in the function setting.
Note: When the AD-8922A-01 is installed, the RS-232C serial interface can not be used.
When connected to the AD-4212C, the both instruments can share power.
For details about the RE-ZERO function of the applicable weighing instruments, refer to
RE-ZERO of Table 2 in "1-2 Applicable Instruments".
10
-
1
BCD Output Specifications
Accessories
I/O plug applicable to the BCD output port
1 pc.
Instruction manual
1 copy
Cable to connect to a weighing instrument (Length: Approx. 2 m) 1 pc.
D-Sub 25-pin to DIN 7-pin (AX-KO577A-200)
Note: A substitute cable may be provided as specified when ordering the AD-8922A.
D-Sub 9-pin to DIN 7-pin (AX-KO1786-200)
DIN 7-pin to DIN 7-pin (AX-KO507-W200)
A round waterproof 12-pin to DIN 7-pin (AX-KO3705-500)
The communications cable provided for the AD-4212C (5m)
Note: When the AD-8922A-01 is installed, the communications cable provided with the
AD-8922A will not be used.
Panel view
BCD output port (BCD-OUT)
AD8922A BCD OUT
BALANCE/SCALE