30
12
CURRENT LOOP INPUT (AD-8922A-05)
Can receive the current loop output from the weighing instrument. The data received can be output to an
external device such as a personal computer and a printer, using the RS-232C serial interface. The
weighing instrument can not be operated using the AD-8922A keys.
Note: When AD-8922A-05 is installed, the pin assignment (DIN 8-pin) of the RS-232C serial
interface, to connect an external device, will be changed.
Note: The AD-4212C does not have current loop output. Therefore, this option can not be used.
12
-
1
Current Loop Input Specifications
Accessories
Cable to connect to a weighing instrument (AX-KO1786-200: Length: Approx.2 m )
DIN 7-pin to D-Sub 9-pin
Instruction manual
1 copy
Note: When the AD-8922A-05 is installed, the communications cable provided with the AD-8922A
as standard will not be used.
Panel view
Current loop input / PC/PRINTER output
Transmission system
: Input: 20 mA current loop (Active) D-Sub 9-pin (Current loop)
Output: EIA RS-232C DIN 8-pin (PC/PRINTER)
Transmission form
: Asynchronous, uni-directional
Data format
: Baud rate
: 600, 1200, 2400, 4800, 9600, 19200 bps
Data bits
: 7 bits or 8 bits
Parity bit
: EVEN, ODD
(Data bits 7 bits)
NONE
(Data bits 8 bits)
Stop bits
: 1 bit or 2 bits
Code
:
ASCII
Note: When a baud rate of 4800 bps or higher is used, communication may not be performed
properly.
AD8922A
OP-05
CURRENT
LOOP
PC/PRINTER
Содержание AD-8922A
Страница 1: ...AD 8922A Remote Controller INSTRUCTION MANUAL 1WMPD4002022 ...
Страница 18: ...16 8 EXTERNAL DIMENSIONS Unit mm 238 130 132 57 170 192 185 96 91 108 72 100 ...
Страница 40: ...38 ...
Страница 41: ...39 ...
Страница 42: ...40 ...