AD-8922A 

 

 

 

Remote Controller 

INSTRUCTION MANUAL 

1WMPD4002022 

Содержание AD-8922A

Страница 1: ...AD 8922A Remote Controller INSTRUCTION MANUAL 1WMPD4002022 ...

Страница 2: ...oduced transmitted transcribed or translated into any language in any form by any means without the written permission of A D Company Ltd The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice ...

Страница 3: ... 15 8 EXTERNAL DIMENSIONS 16 9 OPTIONS 17 9 1 Confirming the AD 8922A Software Version 17 9 2 Installing the Option 18 10 BCD OUTPUT AD 8922A 01 19 10 1 BCD Output Specifications 19 10 2 Setting of the Decimal Point Position 24 11 COMPARATOR OUTPUT AD 8922A 04 25 11 1 Comparator Output Specifications 25 11 2 Using the Comparator Output 27 11 3 Setting the Upper and Lower Limit Values 28 12 CURRENT...

Страница 4: ... both instruments can share power Power can be supplied to both instruments by plugging the AC adapter into either the AD 4212C or the AD 8922A For details refer to 3 3 Turning the Power on 1 2 Applicable Instruments Applicable weighing instruments and what is required are shown in the table below Table 1 Applicable weighing instruments and what is required What is required to connect to a weighin...

Страница 5: ... in the table indicates that the key operation is not available 1 Switching the standby or weighing mode is available for the AD 4212C 2 Not applicable to the counting mode and the percent mode Switching the minimum display is not available for the ET W 3 Available when the AD 8922A is in key mode out 1 or out 2 of the function setting 4 Not available for the ET W and AD 4212 5 AD 8922A function s...

Страница 6: ...X KO1786 200 When connecting to the AD 4212C use the cable provided for the AD 4212C Do not use the cable provided for the AD 8922A Rear Connector to the weighing instrument D Sub 9 pin Connector to the external device Display Keys Angle adjustment knob AC adapter jack Grounding terminal Accessories AC adapter AC adapter identification label See Note below Communications cable to connect to a weig...

Страница 7: ...00 g the display is like 2 2 Key Operation Available key operations to control the weighing instrument depend on the weighing instrument used For details see Table 2 in 1 2 Applicable Instruments Set the command set C5et of the function setting appropriate to the weighing instrument To enter the function setting mode of the AD 8922A press and hold the ON OFF key and press the CAL key The weighing ...

Страница 8: ...se 2 Set the command set C5et of the function setting appropriate to the weighing instrument See Table 2 in 1 2 Applicable Instruments Note Some items may not be available for the instrument used For details on setting refer to the instruction manual of each weighing instrument When connected to an external device such as a personal computer or a printer set the output mode out and time out Hold o...

Страница 9: ... 1 2 Applicable Instruments 3 5 Example of Use A personal computer is connected to the AD 8922A and the weighing data can be transmitted to the personal computer using Windows communication tools WinCT To protect against inadvertent operations the AD 8922A keys can be disabled Function setting C5et 0 When nothing has been received for two seconds or more the bar display appears By changing the fun...

Страница 10: ... 8121B MODE 1 Prints data when the AD 8121B DATA key is pressed Monitors the weighing data of the weighing instrument remotely Key mode or Autoprintmode Outputs the weighing data when the key is pressed or outputsthedata automatically when stable Through mode out 0 Sends all of the received weighing data to the external device AD 8121B MODE 1 Prints data according to the data output mode of the we...

Страница 11: ...e sets 9600 bps for Baud rate and 8 bits NONE for Data bits parity bit Note Use much care when changing parameters The AD 8922A may not function properly when the settings and operational environment are not appropriate 4 1 Display and Keys Table 5 Display Key Description The symbol indicates that the parameter displayed is in effect Selects the class or item in the function setting mode Changes t...

Страница 12: ...eout 1 Hold display Displays the previous data received if nothing has been received for two seconds or more and turns on the hold indicator 0 Not used No buzzer upon data receipt 1 At intervals Sounds buzzer when the data is received with an interval of two seconds or more bep Data receipt buzzer 2 Used Sounds buzzer upon each data receipt Not set 0 dpp Position of decimal point 6 Set Sets the de...

Страница 13: ... value is stable or overloaded 2 Comparison including near zero when the value is stable or overloaded 3 Continuous comparison excluding near zero 4 Continuous comparison including near zero Cp Comparator mode 5 Contact outputs the second header information of the data received Applicable toAD 4212A GP and GX K 0 OFF bep LO buzzer 1 ON Selects whether or not to sound the buzzer when LO 0 OFF bep O...

Страница 14: ...e weighing data display appears 2 While pressing and holding the ON OFF key press the PRINT key Clr appears in the display 3 Press the PRINT key To cancel the operation press the CAL key 4 Press the RE ZERO key to select go 5 Press the PRINT key to perform initialization After initialization the bar display or the weighing data display appears ...

Страница 15: ...ata other than weighing data is output The bar display remains and no weighing data appears Is the weighing instrument in the stream mode Only the stream mode displays the weighing data continuously Other modes display the data only when received When the hold display is selected in the AD 8922A function setting the previous data received is displayed until the next data is received Are the commun...

Страница 16: ...d rate 600 1200 2400 4800 9600 19200 bps Refresh rate Approx 10 times second When Baud rate is 2400 bps or greater Applicable connectors D Sub 9 pin Male to connect to the weighing instrument DIN 8 pin Female to connect to an external device Dimensions 238 W x 132 D x 170 H mm Mass Approx 1 0 kg Standard accessories AC adapter Communications cable Approx 2 m Please confirm that the AC adapter type...

Страница 17: ...t Receive data 3 TXD Output Transmit data 3 TXD Output Transmit data 4 N C 4 RTS Input Request to send 5 SG Signal ground 5 SG Signal ground 6 DSR Input Data set ready 6 CTS Output Clear to send 7 RTS Output Request to send 7 DSR Output Data set ready 8 CTS Input Clear to send 8 N C 9 Va Used internally The signal names except TXD and RXD apply to the DTE such as a personal computer AD 8922A is a ...

Страница 18: ...16 8 EXTERNAL DIMENSIONS Unit mm 238 130 132 57 170 192 185 96 91 108 72 100 ...

Страница 19: ...ents can share power AD 8922A 05 Current loop input Receives the current loop output from the weighing instrument and displays the weighing data The weighing data received can be output using the RS 232C serial interface The AD 4212C does not have current loop output Therefore this option can not be used AD 8922A 06 Analog output Converts the specified digits of the weighing data received from the...

Страница 20: ...he same for all the options 1 Disconnect the AC adapter 2 Remove the two screws and pull out the RS 232C board provided as standard as shown in the illustration 3 Insert the option board along the guides on the left and right sides 4 Using the two screws removed in step 2 fasten the option board ...

Страница 21: ...ed the RS 232C serial interface can not be used When connected to the AD 4212C the both instruments can share power For details about the RE ZERO function of the applicable weighing instruments refer to RE ZERO of Table 2 in 1 2 Applicable Instruments 10 1 BCD Output Specifications Accessories I O plug applicable to the BCD output port 1 pc Instruction manual 1 copy Cable to connect to a weighing ...

Страница 22: ... 8 103 20 1 21 2 22 4 23 8 104 46 1 47 2 48 4 49 8 105 24 1 All output open collector withstand voltage 30 V no pull up resistor low level output current 48 mA Output logic of data status and strobe can be switched individually in the function table bcd 25 2 Input logic 30 4 Input 31 8 106 32 1 BUSY Data will be held during ON when connected to input signal GND 33 2 34 4 RE ZERO RE ZERO will be pe...

Страница 23: ... Over 0 0 OFF 1 ON Display BCD output The example above is when the output logic has been set at the factory The decimal point information will not be output Balance interior Input signal GND 3 CPU input terminal Use a shielded cable Housing Cable shield When a photocoupler is used RE ZERO 5 Input terminal FG Upon switch ON make the voltage between the input terminal and the input signal GND termi...

Страница 24: ...ZERO input ON is the condition that BUSY is connected to input signal GND Pin 3 The AD 8922A 01 can select either the BUSY input or the RE ZERO input by setting the function When keeping the on state for 100msec the weighing instruments keep the re zero state ON OFF 1 0 ON OFF ON OFF Tstr Tstr 100msec After RE ZERO ON OFF Receive data RS 232C BCD output STROBE BUSY RE ZERO ...

Страница 25: ...e Data bits 8 bits Stop bit 1 or 2 bits Code ASCII Terminator CR or CR LF Circuit DIN 8 pin Pin assignment BALANCE SCALE Pin No Signal name Direction Description 1 Vs Internally used 2 TXD Output Transmit data 3 RXD Input Receive data 4 5 SG Signal ground 6 Va Internally used 7 8 When making the cable yourself do not connect to the internally used terminals SG TXD RXD AD 8922A interior To a weighi...

Страница 26: ...ing the decimal point position dpp Factory setting Keyoperation Balance display AD 8922A display BCD output 00012345 00123456 When changing the minimum display digit by pressing SAMPLE key the BCD shifts the output left and adds the last digit Example 2 When setting the decimal point at the third digit position dpp 3 Keyoperation Balance display AD 8922A display BCD output 00123450 00123456 When c...

Страница 27: ...g 1 pc Instruction manual 1 copy Panel view Comparator output COMP OUT Maximum contact voltage 100 VDC Maximum contact current 100 mA DC Maximum contact resistance 20 Ω Comparator output judgement conditions when upper limit value lower limit value Weighing data upper limit value Activates the HI comparator output Upper limit value weighing data lower limit value Activates the OK comparator output...

Страница 28: ... 1 N C 1 HI HI contact output 2 RXD Input Receive data 2 COM COM contact output 3 TXD Output Transmit data 3 TXD Transmit data RS 232C output 4 N C 4 LO LO contact output 5 SG Signal ground 5 SG Signal ground 6 DSR Input Data set ready 6 OK OK contact output 7 RTS Output Request to send 7 DSR Data set ready RS 232C output 8 N C 8 RXD Receive data RS 232C output 9 N C AD 8922A is a DTE Connects to ...

Страница 29: ...in the Buzzer mode bep of the Comparator Cp fnc Note When setting the upper and lower limit values make sure that the upper limit value is greater than the lower limit value Function setting The function setting Cp fnc is available only when the AD 8922A 04 is installed Class Item Parameter Description 0 No comparison 1 Comparison excluding near zero when the value is stable or overloaded 2 Compar...

Страница 30: ... Displaying example 2 Select the comparator number using the following keys RE ZERO key To increase the comparator number by 1 MODE key To decrease the comparator number by 1 Each time the key is pressed the upper limit value and the lower limit value of the comparator number selected is displayed alternately C01 HI C01 LO C02 HI C02 LO 3 Press the SAMPLE key to go to the storing mode to change th...

Страница 31: ...e comparator number by 1 Each time the key is pressed the upper limit value and the lower limit value of the comparator number selected is displayed alternately C01 HI C01 LO C02 HI C02 LO Only the stored comparator numbers are displayed 4 Press the PRINT key to confirm the selection and return to the weighing data display with the selected upper and lower limit values ready for use In the example...

Страница 32: ...ssories Cable to connect to a weighing instrument AX KO1786 200 Length Approx 2 m DIN 7 pin to D Sub 9 pin Instruction manual 1 copy Note When the AD 8922A 05 is installed the communications cable provided with the AD 8922A as standard will not be used Panel view Current loop input PC PRINTER output Transmission system Input 20 mA current loop Active D Sub 9 pin Current loop Output EIA RS 232C DIN...

Страница 33: ... 2 C L Current loop 2 N C 3 N C 3 TXD Transmit data RS 232C output 4 N C 4 N C 5 C L Current loop 5 SG Signal ground 6 N C 6 N C 7 N C 7 DSR Data set ready RS 232C output 8 N C 8 N C 9 N C 9 8 7 6 SG DSR AD 8922A interior TXD 6 1 4 2 5 3 7 8 5 4 3 2 1 C L C L 5V 3 3 k Ω To a weighing instrument To an external device such as a personal computer or a printer Inch screw thread 4 40UNC AD 8922A interi...

Страница 34: ...s DIN connector Plug 1 pc Screwdriver 1 pc Instruction manual 1 copy Panel view Analog output Output impedance 100 Ω or less Linearity 0 3 or less Output range 0 V 1 V With the slide switch set to 0V 0 2 V 1 V With the slide switch set to 0 2V RS 232C Transmission system EIA RS 232C Transmission form Asynchronous bi directional half duplex Data format Baud rate 600 1200 2400 4800 9600 19200 bps Da...

Страница 35: ...ata 4 N C 4 N C 5 SG Signal ground 5 SG Signal ground 6 DSR Input Data set ready 6 DSR Output Data set ready 7 N C 7 AOUT Output Analog output 8 N C 8 RXD Input Receive data 9 N C AD 8922A is a DTE Connects to a DCE such as a weighing instrument using a straight through cable 9 8 7 6 SG 10 Ω SG AD 8922A interior AD 8922A interior TXD 6 1 4 2 5 3 7 8 5 4 3 2 1 Analog GND RXD Analog output D S R TXD...

Страница 36: ...utput Converts the consecutive 3 digits with the digit selected in 5el as the least to voltage and outputs 0 Selects the first digit as the least 1 Selects the second digit as the least 2 Selects the third digit as the least 3 Selects the fourth digit as the least 4 Selects the fifth digit as the least aout Analog output 5el Analog output digit selection 5 Selects the sixth digit as the least Fact...

Страница 37: ...put voltage can be fine adjusted 13 4 1 Fine adjustment procedure 1 While pressing and holding the ON OFF key press the RE ZERO key At this time the output voltage will be at zero 2 Turn the ZERO knob so that the voltmeter indicates 0 000 V when the slide switch is set to 0V 0 200 V when the slide switch is set to 0 2V 3 Press the RE ZERO key At this time a voltage of 1 V is generated Turn the SPA...

Страница 38: ... is fixed under the following conditions 1 While the weighing data is not displayed 0 V or 0 2 V e g the bar display function setting operation 2 When e Weighing pan error is being displayed 0 V or 0 2 V 3 When e Overload error is being displayed 1V ...

Страница 39: ...A 200 EK i EW i FC i FC Si GH HR i FZ i FX i None D Sub 9 pin AX KO2466 200 AX KO1786 200 EK G EK H ET W EW G OP 03 D Sub 25 pin AX KO1710 200 AX KO577A 200 HV G HV WP HW G HW WP None DIN 7 pin AX KO1786 200 AX KO507A 200 FG OP 03 DIN 7 pin AX KO1786 200 AX KO507 W200 FS FS KL OP 03 DIN 8 pin AX KO1786 200 AX KO507 W200 FG L FG M OP 23 DIN 8 pin AX KO1786 200 AX KO507 W200 Note A substitute commun...

Страница 40: ...38 ...

Страница 41: ...39 ...

Страница 42: ...40 ...