Page
39
AD-4408A
10.2.4. Checking interfaces
Refer to the instruction manual for each interface, available on our website.
10.2.5. Checking the testing terminal
The
Chc r5
display is for checking the testing terminal. Do not use it.
10.2.6. Checking the A/D (load cell)
With
Chc ad
displayed, press the ENTER key to enter the A/D check mode. The load cell
output value is displayed in mV/V.
The example below is when the internal count is 1.23456 mV/V.
10.2.7. Checking the internal count
With
Chc in
displayed, press the ENTER key to enter the internal count check mode.
The internal count value (display value times 10) is displayed.
The example below is when the internal count is 123.
10.2.8. Checking the program version
With
Chcpr9
displayed, press the ENTER key to display the version of the current
program. The example below is when the program version is 1.00.
10.2.9. Checking the serial number
With
Chc 5n
displayed, press the ENTER key to display the last five digits of the serial
number. The example below is when the last five digits of the serial number are 12345.
10.2.10. Checking the program checksum
With
C5 pr9
displayed, press the ENTER key to display the program checksum. The
example below is when the checksum is EF.
1 . 2 3 4 5 6
1 2 3 .
p
0 0
1 . 0 0
C 5
0 0
e f
1 2 3 4 5 .