AD-4408A
Page
12
5.1.2. Power source
Before connecting the indicator to the power source, read the instruction manual
thoroughly.
Do not connect the indicator to the power source before the installation is complete.
To avoid electrical shock, do not handle the power cable with a wet hand.
Earth ground the indicator. Do not share the ground line with other electrical power
equipment.
The power requirement is 100 VAC to 240 VAC and the frequency is 50 Hz or 60 Hz.
Use a stable power source free from instantaneous power failure or noise.
To avoid a malfunction, do not share the power line with other devices.
The output voltage of a load cell is a very sensitive signal. Keep all electrical noise
sources away from the load cell and load cell cable.
Use shielded I/O cables. Connect the cable shield to the F.G. terminal or the
indicator housing.
Earth ground the indicator to prevent electrical shock or
indicator malfunction.