AD-4408A
Page
10
4.2. Rear Panel
Fig.2 Rear panel
1 Terminal block
Terminal No.
Function
1
EXC+
Load cell excitation voltage (+)
2
SEN+
Sensing input (+)
3
SEN-
Sensing input (-)
4
EXC-
Load cell excitation voltage (-)
5
SIG+
Load cell input (+)
6
SIG-
Load cell input (-)
7 SHLD
Shield
Load cell
connection
8 C.L.
Standard
serial
output (Current loop)
9 C.L.
Standard
serial
output (Current loop)
10 F.G.
Frame
ground
11 E Earth
ground
12 AC AC
power
source
13 AC AC
power
source
NOTE: Terminals 7 (SHLD) and 11 (E) are internally connected.
To avoid unintentional operations, keep the cover attached on the terminal block.
2 Blank panel
Option slot inside to install the interface module.
3 Testing terminal
Do not use, leave this terminal unconnected.
4 Sealing bolts
Can be sealed with a wire seal.
2 3
1
4