Page
23
AD-4408A
Function No.
Parameter or
Setting range
Function Description
Default
value
Calf01
0
to
4
Unit
Unit of the weight value
0
: None
1
: g
2
: kg
3
: t
4
: lb (USA version)
2
Calf02
0
to
5
Decimal
point
position
Decimal point position of the weight value
0
: None 123456
1
: 10
1
12345.6
2
: 10
2
1234.56
3
: 10
3
123.456
4
: 10
4
12.3456
5
: 10
5
1.23456
0
Calf03
1
to
6
Minimum
division
Minimum division (d) of the weight value
1
: 1
2
: 2
3
: 5
4
: 10
5
: 20
6
: 50
1
Calf04
1
to
999999
Weighing
capacity
Weighing capacity of the weighing
instrument
Weighing is possible up to the value of this
setting plus 8 d. If the value exceeds this,
overflow will occur and will not be
displayed.
The decimal point position is the same as
the setting of
Calf02
.
20000
Calf05
0
to
100
Zero
adjustment
range
Range to enable zero adjustment by the
ZERO key
Expressed in percent of the weighing
capacity with the calibration zero point as
the center.
For example, if this is set to 2, the value in
the range of
±
2% of the weighing capacity
with the calibration zero point at the center
will set to zero.
2