AD-4408A
Weighing Indicator
for Field Networks
INSTRUCTION MANUAL
1WMPD4001973
Страница 1: ...AD 4408A Weighing Indicator for Field Networks INSTRUCTION MANUAL 1WMPD4001973...
Страница 2: ...laced This symbol indicates the ground terminal This symbol indicates that an operation is prohibited NOTE Information or cautions to use the device correctly 2009 A D Company Limited All rights reser...
Страница 3: ...5 1 1 Environment 11 5 1 2 Power source 12 5 2 Load Cell Connection 13 5 3 Interface Module Installation 14 6 Operation 15 6 1 General Function 15 6 1 1 Zero adjustment 15 6 1 2 Zero tracking 15 6 1...
Страница 4: ...ces 39 10 2 5 Checking the testing terminal 39 10 2 6 Checking the A D load cell 39 10 2 7 Checking the internal count 39 10 2 8 Checking the program version 39 10 2 9 Checking the serial number 39 10...
Страница 5: ...protect against a short circuit and possible fire hazard The indicator is equipped with various protection circuits and the fuse should not burn out under normal operations If the fuse burns out the...
Страница 6: ...CompactCom network interface module manufactured by HMS Industrial Networks The interface module is available as an accessory for the AD 4408A Purchase the interface module from A D or directly from a...
Страница 7: ...ethod Delta sigma method A D resolution count Approximately 16 000 000 counts Display resolution 999 999 d max Less than 20 000 d recommended Sampling rate 100 times second 3 2 Digital Section Display...
Страница 8: ...ity Zero tracking Tracks the weight drift around the zero point to maintain zero Zero tracking time 0 0 to 5 0 seconds Can be set optionally within the range Zero tracking band 0 0 to 9 9 d Can be set...
Страница 9: ...ource 100 VAC to 240 VAC 10 15 50 60 Hz Power consumption Approximately 10 VA Operating temperature 10 C to 40 C Operating humidity 85 R H or less no condensation Installation method Panel mount Mass...
Страница 10: ...ns 4 GROSS Turns ON when the displayed value is a gross weight 5 NET Turns ON when the displayed value is a net weight 6 HOLD Turns ON when the weight value is being held Two hold modes can be selecte...
Страница 11: ...print command Hold Alternate switch Momentary switch Clear the tare weight Clear the zero value The key decreases the value of the blinking digit by one when the numerical values are being entered 13...
Страница 12: ...Standard serial output Current loop 9 C L Standard serial output Current loop 10 F G Frame ground 11 E Earth ground 12 AC AC power source 13 AC AC power source NOTE Terminals 7 SHLD and 11 E are inte...
Страница 13: ...When used in an environment where the indicator may be exposed to splashing water secure the front panel to the control panel using the accessory panel mount packing It makes the front panel water res...
Страница 14: ...ical power equipment The power requirement is 100 VAC to 240 VAC and the frequency is 50 Hz or 60 Hz Use a stable power source free from instantaneous power failure or noise To avoid a malfunction do...
Страница 15: ...long or when summing multiple load cells to avoid drift caused by changes in temperature If the cable is short a 4 wire shielded cable with terminals 1 and 2 EXC and SEN and terminals 3 and 4 SEN and...
Страница 16: ...the PC board connector located in the option slot Step 4 Using a TORX driver size T9 fasten the retaining screws with a tightening torque of 0 25 Nm in the clockwise direction to secure the interface...
Страница 17: ...he indicator while holding the TARE key or with the indicator in the OFF mode press and hold the TARE key and press the ON OFF key to clear the zero value Functions related to zero adjustment z Calf05...
Страница 18: ...s a function to the F key from the functions below 0 None 1 Manual print command 2 Hold 3 Alternate switch Read the description below 4 Momentary switch Read the description below 5 Clear the tare wei...
Страница 19: ...range setting error over 3 F key status Read the description below F key status zEnables when alternate switch or momentary switch is selected for fncf02 F key The display turns ON when the F key is...
Страница 20: ...sconnected while in the OFF mode the indicator will start in the OFF mode when power is re connected Fig 6 Mode map Power ON OFF mode Weighing mode General function mode Maintenance mode Calibration m...
Страница 21: ...n related function Setting basic constant values of the indicator such as the minimum division and weighing capacity and other data directly related to weighing is performed Setting the parameters for...
Страница 22: ...zero calibration is not to be performed press the F key and go to step 5 To monitor the current weight value press the CAL key Press the CAL key again to display CalL 0 Step 4 With nothing placed on...
Страница 23: ...ibration data is written into the FRAM Step 9 The current state is the same as that of step 2 To go back to the weighing mode press the ESC key again NOTE The blinking decimal point indicates that the...
Страница 24: ...e are available depending on the function number as below Method Description Selecting a parameter Only the available parameter is displayed and blinks Press the or key to select a parameter Inputting...
Страница 25: ...5 20 6 50 1 Calf04 1 to 999999 Weighing capacity Weighing capacity of the weighing instrument Weighing is possible up to the value of this setting plus 8 d If the value exceeds this overflow will occ...
Страница 26: ...acking using this setting in combination with the setting of the zero tracking time When 0 0 zero tracking will not be performed Unit d minimum division in 0 1 increments 0 0 Zero tracking follows the...
Страница 27: ...ility detection will not be performed Stable all the time Unit d minimum division 2 Calf10 0 to 1 Tare and zero adjustment when unstable Tare and zero adjustment when the weight value is unstable 0 Di...
Страница 28: ...bles 1 Calf16 0 to 1 Zero upon power ON Select whether or not to perform zero upon power ON 0 Disables 1 Enables 0 Calf17 7 00000 to 7 00000 Input voltage at zero Input voltage from a load cell at zer...
Страница 29: ...n actual load Calf20 0 to 1 Output when the gross weight is negative Output by the manual print command when the gross weight is negative 0 Enables output 1 Disables output 0 Calf21 0 to 1 Communicati...
Страница 30: ...ck the load cell rating and connection When nothing is wrong with the rating and connection adjust the load cell output as described in the next section When the load cell or A D converter may be the...
Страница 31: ...output Use a resistor with a high resistance value and a low temperature coefficient Fig 7 Load cell output adjustment C Err 2 When exceeding in the positive direction C Err 3 When exceeding in the n...
Страница 32: ...ame fncf Basic function Cl f Standard serial output f f f Interfaces Refer to the instruction manual for each interface available on our website Press the ENTER key The function number will be display...
Страница 33: ...to the state of step 3 Press the ESC key to store the setting values in the FRAM and go back to the weighing mode NOTE The blinking decimal point indicates that the current value is not the weight val...
Страница 34: ...ch 4 Momentary switch 5 Clear the tare weight 6 Clear the zero value NOTE Clearing the zero value can be enabled or disabled by Calf15 0 fncf03 1 to 3 Display update rate 1 20 times second 2 10 times...
Страница 35: ...o 3 Data transfer mode 1 Stream 2 Auto printing Note 1 3 Manual printing 1 Cl f03 1 to 2 Baud rate 1 600 bps 2 2400 bps 2 Cl f04 0 to 99 ID number Adds an ID number to the standard serial output Note...
Страница 36: ...d in the general functions Cl f01 to Cl f04 Type 0 to 20 mA current loop Data bits 7 bits Start bit 1 bit Parity 1 bit even Stop bit 1 bit Baud rate 600 2400 bps Code ASCII 9 1 1 Connection Fig 8 Stan...
Страница 37: ...t examples Gross weight Net weight Tare weight Data with a decimal point Positive overflow Negative overflow Unstable data Output OFF data NOTE The decimal point position is the same even when overflo...
Страница 38: ...a that is not displayed is never transmitted Auto printing mode Transmits data once only when the weight value with 5 d or greater is stable For another transmission the weight value must fall below 5...
Страница 39: ...rrors err dt The setting value is out of the settable range Check the setting value and set again 10 2 Check Mode The check mode checks the performance of the display key switches and external I O 10...
Страница 40: ...press the ENTER key to enter the key switch check mode In the key switch check mode six squares are displayed Each square corresponds to a key switch as shown below When a key switch is pressed the c...
Страница 41: ...the internal count check mode The internal count value display value times 10 is displayed The example below is when the internal count is 123 10 2 8 Checking the program version With Chcpr9 displayed...
Страница 42: ...t counted The example below is when the checksum is EF 10 2 12 Checking the calibration related functions With CalfdtP displayed press the ENTER key to display the settings of the calibration related...
Страница 43: ...ons initialization Step 1 While pressing and holding the ENTER key press the F key fnc is displayed to indicate that the indicator will enter the general function mode To go back to the weighing mode...
Страница 44: ...l load To go back to the weighing mode press the ESC key Step 3 Press the or key to select the all data initialization mode iniallLL and press the ENTER key Step 4 All the status LEDs blink to prompt...
Страница 45: ...Assigns a function to the F key 0 None 1 Manual print command 2 Hold 3 Alternate switch 4 Momentary switch 5 Clear the tare weight 6 Clear the zero value 0 fncf03 1 to 3 Display update rate 1 20 times...
Страница 46: ...ue User setting Cl f01 1 to 5 Output data 1 Displayed weight 2 Gross weight 3 Net weight 4 Tare weight 5 Gross Net Tare 1 Cl f02 1 to 3 Data transfer mode 1 Stream 2 Auto printing 3 Manual printing 1...
Страница 47: ...apacity with the calibration zero point as the center 2 Calf06 0 0 to 5 0 Zero tracking time Unit second When 0 0 zero tracking will not be performed 0 0 Calf07 0 0 to 9 9 Zero tracking band Unit 0 1...
Страница 48: ...1 Select whether or not to clear the zero value 0 Disables 1 Enables 1 Calf16 0 to 1 Select whether or not to perform zero upon power ON 0 Disables 1 Enables 0 Calf17 7 00000 to 7 00000 Input voltage...
Страница 49: ...Page 47 AD 4408A 12 External Dimensions Unit mm Fig 10 External dimensions 144 72 134 7 6 10 99 5 16 3 2 138 1 0 68 0 7 0 Panel cutout dimensions Net B G...
Страница 50: ...AD 4408A Page 48 MEMO...