7
ATTACHING THE AD-4212C SERIES WEIGHING UNIT DIRECTLY ON THE MOUNTING BASE
When the AD-4212C series weighing unit is built into a system, remove the three leveling feet and
use the screw holes to secure the weighing unit to the base.
Attachment Procedure
1. Remove the three leveling feet.
Caution
Do not remove any other screws from the
bottom of the weighing unit. Doing so could
damage the weighing sensor.
2. Secure the weighing unit from above.
(Prepare appropriate screws with a size
corresponding to an M5 screw. The screw
hole diameter: 6 mm).
In case of securing the weighing unit from
the under side, use M6 screws with 1-mm
pitch.
Notes
The screws to secure the weighing unit
to the mounting base are not provided.
For the position of the holes, refer to “10.
EXTERNAL DIMENSIONS”.
When the leveling adjustment is difficult to perform due to the installation conditions, place a
shim between the lower surface of the weighing unit and the securing surface, or use two nuts. If
the adjustment is still difficult to perform, perform calibration before use. Then, the balance will
function normally.
To secure
Leveling foot
Weighing
unit