20
6. COMMUNICATION SPECIFICATIONS
The AD-4212C series can communicate interactively using RS-232C.
The AD-4212C series can continuously transmit the weighing data. (Approx.10 times/second
*
to 50
times/second)
The AD-4212C series is DCE. The AD-4212C series can connect directly to the RS-232C terminal of a
personal computer by using the accessory RS-232C cable.
*
Factory setting
1. RS-232C specification
Transmission system
EIA RS-232C
Transmission form
Asynchronous, bi-directional, half duplex
Transmission rate
Approx.10 times/second
*
to 50 times/second
Data format
Baud rate
2400bps
*
to 19200bps
Data bits
7 bits
Parity
EVEN
Stop bit
1 bits
Code
ASCII
Terminator
<CR><LF>
*
Factory setting
Relation between baud rate and transmission rate of weighing data
Baud rate
Transmission rate of weighing data
2400bps 12.5
times/second
4800bps 25.0
times/second
9600bps 50.0
times/second
19200bps 50.0
times/second
1
0
-5V
to -
15V
+5V
to
+15V
Data bits
Start bit
Parity bit
Stop bit
LSB MSB