![A&D AD-4212C-300 Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/aandd/ad-4212c-300/ad-4212c-300_instruction-manual_3083527030.webp)
28
9. SPECIFICATIONS
AD-4212C-300 AD-4212C-600
AD-4212C-301
Weighing capacity
320 g
620 g
320 g / 51 g
*1
Maximum display
320.084 g
620.084 g
320.084 g / 51.0009 g
Minimum weighing value
(1 digit)
0.001 g
0.001 g
0.001 g / 0.0001 g
Repeatability
(Standard deviation)
0.001 g
0.001 g
0.001 g / 0.0002 g
Linearity
±
0.002 g
±
0.005 g
±
0.002 g
0-30 g
0.5 s
0-30 g
0.5 s
When the minimum
weighing value of
0.0001 g is selected
1.3 s
*2
Stabilization time in seconds
(typical at FAST
under good environment)
30-320 g
1.0 s
30-620 g
1.0 s
When the minimum
weighing value of
0.001 g is selected
1.0 s
Display refresh rate
Approx.10 times/second
*
to 50 times/second
I/O unit
RS-232C
Bi-directional, 2400 bps
*
to 19200 bps
(WinCT-AD4212C communication software provided)
Sensitivity drift
±2 ppm/°C (10°C to 30°C/50
°
F to 86
°
F )
Operating environment
5°C to 40°C (41°F to 104°F), 85% RH or less (No condensation)
Allowable eccentric load
0.1 Nm or less (With the center of the weighing pan as a reference)
Calibration weight provided
200g (Conforming to OIML Class E2)
Applicable weight values
50 g, 100 g,
200 g
*
,
300g
50 g, 100 g,
200 g
*
,
300g,
400 g, 500 g,
600 g
50 g, 100 g,
200 g
*
,
300g
Dimensions 59
(W)
×
231 (D)
×
91 (H) mm
Weighing pan
50
×
50 mm
Weighing
unit
Mass Approx.
1.6
kg
Connection cable
Approx. 10 m
Unit display mode
G
g
AC adapter
Output voltage 12V (Confirm that the adapter type is correct for the
local voltage and power receptacle type)
Power consumption
Approx. 10VA (supplied to the AC adapter)
*1
Smart range (Refer to “5. SMART RANGE FUNCTION”.)
*2
When the precision range is used
*
Factory setting