JANUARY 2010 - TD 220a DEPLOYMENT GUIDE, RDCP600
5.6
Sensor update of Sensor Board
1.
NOTE! This step is only required if Startup Mode is set to:
Start the Recorder when powered up, or
Enable for Remote start.
If Startup Mode is set to
Start Recording from touch panel
, proceed to step 2.
Tap the
Menu
button in the lower left corner of the LCD display to activate the menu.
In the RDCP 600 Menu list tap
Deployment Setup.
In the Deployment Setup Configuration, select the
Edit
menu and then select
Startup
Mode.
Use
Start Recording from touch panel.
Tap OK to return to the Deployment Setup Configuration Application and close the
application (tap the
x
in the upper right corner) to save the Startup Mode settings.
NOTE! Remember to set the Startup Mode back to the desired mode after the sensor
board software is updated.
2.
Switch
OFF
the instrument. Remove the Compact Flash card if connected. Open the
connector cover on top of sensor board cover to access the sensor board. Move jumper 1
one step to the right,, refer to 4-7.
3.
Start the instrument by pressing the power button at the rear of the instrument. Wait for
approximately 10 seconds before the LCD display at the front of the instrument is
activated.
4.
Tap the
Menu
button in the lower left corner of the
LCD display to activate the menu.
In the RDCP 600 Menu list tap
Diagnostics
.
In the Diagnostics menu tap
SB_FW_Inst
(SensorBoard Firmware Installer).