16.
Clean the scars with Acetone.
17.
Apply Repair Lacquer to repair the scars.
18.
Control that the Electrical terminal is securely tightened.
5.1.2
Replacements of Parts
The
Profiler
, the
Turbidity Sensor
and the
Oxygen Optode
are calibrated separately and can be
changed without influencing the instrument performance. The Calibration of the Temperature
sensor, Turbidity sensor and the oxygen optode is valid regardless of which Aanderaa Instrument
it is installed in. This allows for factory supply of calibrated sensors for installation by the
customer.
The
Pressure Sensor
and the
Conductivity Sensor
must be calibrated when connected to the
instrument for optimum performance. If a new Electronic Board is installed, these sensors
should be re-calibrated before deployment.
An
Electronic Board
can be ordered and installed without any adjustments or re-calibration of
the Profiler, the Turbidity Sensor and the Oxygen Sensor.
5.1.3
Factory Service
Factory service is offered for maintenance, repair or calibration of instrument and parts.
When returning instrument or parts for service, always include the
Instrument Service Order
,
Form No. 135, see our web pages under ‘Support and Training’.
Normal servicing time is four to six weeks, but in special cases the service time can be reduced.
A main overhaul and service is recommended at the factory every two years.
5.1.4
Spare Parts and Accessories
The manufacturer always keeps a stock of spare parts, accessories and consumable parts for
quick delivery. Orders may be placed by fax, telephone or mail.
The Oxygen Optode 3830 is optional, refer to TD200c
RDCP Primer
for information about
calibration and maintenance.