An operation is currently running if a function on the screen is colored in yellow.
Example: Gas inlet is ON
5.3 Display
5.4 Basic setup
Step 1:
Check condenser is clean and condensate removed
Step 2:
Check power supply 3 x 400 V ±10%; 50Hz
Step 3:
Place the material in chamber Close the chamber door.
Step 4:
Put the main switch in position 1
Step 5:
Select the preset program and press
"START"
Step 7:
From this moment on the device runs completely automatically
Step 8:
Press
"STOP"
to stop the device (last step is always endless to keep the
condenser cold and system in vacuum)
Step 9:
Press “Gas inlet – Vacuum brake” to be able to open the door and remove the
material
Step 10:
Start manual program “program 0”
and trigger condenser defrost, condenser
defrost will automatically stop after 10 minutes! (condenser defrost is not necessary also
you can leave the door open and wait overnight to melt the condenser naturally)
Tsh … current shelf temperature
P … current/ set pressure in chamber
Tm …temperature on material probe
Tc … current condenser temperature
Step Nr.:1: duration (10 minutes)
Tsh … current/set shelf temperature
P … current/ set pressure in chamber
Tm … temperature on material probe
Tc… current condenser temperature
Содержание 19116384
Страница 1: ...User Manual January 2020 FREEZE DRYER LIOSMART 100 ...
Страница 7: ...User Manual LIOSMART 100 7 Dimensional drawing ...
Страница 8: ...User Manual LIOSMART 100 8 Dimensional drawing shelves view ...
Страница 9: ...User Manual LIOSMART 100 9 Connections ...
Страница 12: ...User Manual LIOSMART 100 12 Trolley Trolley safety bolt Shelves safety bolt ...
Страница 22: ...User Manual LIOSMART 100 22 ...
Страница 27: ...User Manual LIOSMART 100 27 1 7 2 3 6 9 4 5 8 12 13 14 15 16 17 18 10 11 7 3 Correct ions ...
Страница 29: ...User Manual LIOSMART 100 29 ...
Страница 35: ...User Manual LIOSMART 100 35 11 ATTACHMENTS 11 1 CE declaration of conformity 11 2 Electro diagram ...