25
loosing vacuum power or damage of suction motor. Also use exhaust air hose with 40mm
diameter hose.
Exhaust air temperature increases significantly and discharge drastically too hot for most
plastic piping. Therefore, metal piping or heat-resistant pipe is recommended for at least the
first five (5) to eight(8) feet (1.5 ~ 2.5 meters) from the air exhaust valve on the discharge
side. In addition, this piping MUST be guarded and marked
“DANGER–HOT–DO NOT
TOUCH.”
CAUTION
Do not use hoses such as rubber or PVC hose which are not resistant to
chemicals or disinfectants or insufficient flexibility. Only use PVC flexible
spiral hoses with integrated spiral or hoses of equivalent material
3.7.
Suction inlet hose connection
•
Remove the cover of suction inlet and push-in tubing connector in the inlet hole
•
Connect suction inlet hose (Ø40mm spiral hose) at tubing connector and tighten
with teflon insulation tape
3.8.
Waste outlet hose connection
-
Always use spiral hose with Ø20mm
interior
-
Connect hose to the outlet hole with
clockwise
-
Please be tighten with teflon
-
Always be higher position than
drainage
3.9.
Exhaust air hose connection
•
Turn left exhaust air outlet hole clamp
•
Connect exhaust air outlet hose to the hole (strongly push-in)
•
Turn right exhaust air outlet hole clamp for fasten
•
Always place end of the outlet hose out of building or window to prevent noisy
problem, hot temperature and bad odor
Содержание RAIN2+
Страница 35: ...35 ...
Страница 36: ...36 M R12 rev 02 4th May 2019 ...