64220, Rev. 1/April 2019
3M™ GMI™ PS200-LMP AUTO BUMP/CALIBRATION STATION
CREATE/VIEW CERTIFICATES
6-1
6. CREATE/VIEW CERTIFICATES
6.1. Introduction
Test results are stored in the ABC. These results can be transferred to the USB memory stick for
viewing or printing on a PC/Laptop. The USB memory stick supplied with the ABC must be used.
6.2. Transfer Test Results to the USB Memory Stick
1. Verify that the ABC has completed its warm-up sequence and only one green LED is illuminated,
.
Figure 6-1: ABC ON Indicator
2. Insert USB memory stick into the ABC.
All test results stored in the ABC will be automatically transferred to the USB memory stick. This
will take approximately
15 seconds
, during which time the ABC LED's will flash in various combi-
nations and colors.
3.
Wait
until only one green LED is illuminated, as shown in
.
All test results have now been transferred to the USB memory stick. The ABC is ready to use.
4. Remove the USB memory stick and insert into a PC/Laptop.
5. The typical contents of the USB memory stick are illustrated in
Figure 6-2: USB Memory Stick Contents